3.1 Preparation before Use
3. inspired™ O2FLO Operating
Instruction
15
16
2.5 Specification
Power Input
Power
Flow Rate Range (Adult Mode)
Flow Rate Range (Paediatric Mode)
Oxygen Concentration Accuracy
Working Pressure
Dew Point Temperature Range (Adult Mode)
Dew Point Temperature Range (Paediatric Mode)
Humidification Output
Sound Pressure Level
Dimensions
Net Weight
Expected Service Life of the Unit
100-240V , 50/60Hz
300VA
10~60L/min
2~25L/min
±5%
30~80 psi (0.2Mpa~0.55Mpa)
31~37°C, ±2°C
31~34°C, ±2°C
>33mg/L @ 37°C ; >12mg/L @ 31~36°C
≤50 dB(A)
243mm x 220mm x 170mm
2.8±0.2kg
5 years
1) Install the autofeed humidification chamber by pressing down the chamber lock and
sliding the chamber into position sitting on the heater plate. Once the chamber is in
the correct position, the chamber lock will click into place with the chamber inlet
connecting directly into the annular tube of the machine.
2) Connect water bag
Push the spike at the end of water supply tube into the port at the bottom of the bag.
Open the vent cap on the side of the spike. Hang the water bag above the unit.
The chamber will be automatically filled to the required level and maintain at that level.
Note:
The unit will beep and the status LED will flash in yellow if there is insufficient
water in the chamber. In this situation, check and make sure that the water supply tube
is free of kinks and that there is sufficient water in the water bag. Replace as necessary if
a kink is found or the water bag is empty.
L/min
hrs
2
340
5
136
10
68
15
46
20
34
25
27
30
22
35
19
40
17
45
15
50
14
55
12
60
11
VHC60: Flow setting vs Typical usage time (2 liter sterile water bag).