Guidance and manufacturer’s declaration – electromagnetic immunity

a) The interface between the PATIENT physiological signal simulation, if used, and the ME 

EQUIPMENT or ME SYSTEM shall be located within 0,1 m of the vertical plane of the uniform 

field area in one orientation of the ME EQUIPMENT or ME SYSTEM.
b) ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for 

the purpose of their operation shall be tested at the frequency of reception. Testing may be 

performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS. 

This test assesses the BASIC SAFETY and ESSENTIAL PERFORMANCE of an intentional receiver 

when an ambient signal is in the passband. It is understood that the receiver might not 

achieve normal reception during the test.
c) Testing may be performed at other modulation frequencies identified by the RISK 

MANAGEMENT PROCESS.
d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components 

or circuitry.
e) During the test, the ME EQUIPMENT or ME SYSTEM may be powered at any NOMINAL 

input voltage, but with the same frequency as the test signal. (see Table 1)
f) Before modulation is applied.
g) This test level assumes a minimum distance between the ME EQUIPMENT or ME SYSTEM 

and sources of power frequency magnetic field of at least 15 cm. If the RISK ANAYLSIS 

shows that the ME EQUIPMENT or ME SYSTEM will be used closer than 15 cm to sources of 

power frequency magnetic field, the IMMUNITY TEST LEVEL shall be adjusted as appropriate 

for the minimum expected distance.

Guidance and manufacturer’s declaration – electromagnetic emissions

The VNT275 equipment is intended for use in the electromagnetic environment specified 

below. The customer or the user of the VNT275 should ensure that it is used in such an 

environment.

  Emissions Test

Compliance

Electromagnetic environment – guidance

RF Emissions 

CISPR 11

Group 1

The ME equipment uses RF energy only for its 

internal function. Therefore, its RF emissions 

are very low and are not likely to cause any 

interference in nearby electronic equipment.

RF Emissions 

CISPR 11

Class B

Complies

Harmonic emissions 

IEC 61000-3-2

Not  

Applicable

The ME equipment is  

solely battery powered

Voltage fluctuations/ 

flicker emissions

Not 

Applicable

Содержание A005127R2

Страница 1: ...Owner s Manual REF VNT275US NonContact Infrared Body Thermometer...

Страница 2: ...800 477 0457 Email ConsumerRelations HelenofTroy com Visit www VicksThermometers com Please be sure to specify the model number 2020 All rights reserved Kaz USA Inc a Helen of Troy Company 400 Donald...

Страница 3: ...ontact Infrared Body Thermometer is a medical device and can be used repeatedly for 1 year Please read the manual before using the device if you do not fully understand these instructions for use do n...

Страница 4: ...NOT store this thermometer in temperature below 4 F or over 131 F extremes below 20 C or over 55 C or in excessive humidity up to 93 non condensing relative humidity Let the thermometer and individua...

Страница 5: ...37 weeks of pregnancy are completed Small for gestational age is defined as a newborn baby birth to 4 weeks of age born at 37 weeks or later with a weight below the 10th percentile for newborn babies...

Страница 6: ...the IR Sensor acquires infrared energy from the target forehead Taking into account the ambient temperature the thermometer turns the signal from the IR Sensor to a digital value and displays it on t...

Страница 7: ...ite Axillary Measuring range Body mode 93 2 F 109 2 F 34 0 C 42 9 C Surface mode 32 F 212 F 0 C 100 C Laboratory accuracy Body mode 0 4 F 0 2 C for the range 95 F 107 6 F 35 0 C 42 0 C 0 5 F 0 3 C out...

Страница 8: ...ereas for mercury in glass and electronic thermometers the requirement per ASTM Standards E 667 86 and E 1112 86 is 0 1 C 0 2 F This device conforms to the following standards IEC 60601 1 Medical elec...

Страница 9: ...ot required for extended periods of time in order to avoid damage to the thermometer resulting from a leaking battery 2 Operation 2 1 Battery installation CAUTION The Non Contact Infrared Body Thermom...

Страница 10: ...direct sunlight 3 General Setup and Use Start measuring Align the device to forehead at a distance of 1 5cm 0 4 2 0 inches then press the On Trigger button to start the measurement and read the data T...

Страница 11: ...nded to keep this thermometer at the settings originally programed on the device Fever Alert Set F2 This setting will change at what point the backlight color of the temperature reading turns to red A...

Страница 12: ...cker automatically turns off Replace battery or the product has been damaged and needs repairs Low Battery Warning When the battery is running low the display flashes the low battery warning symbol ho...

Страница 13: ...Storage The lens is very delicate It is very important to protect the lens from dirt and damage Use a clean soft cloth to clean the surface of the device and LCD Do not use solvents or immerse the de...

Страница 14: ...mitations or exclusions may not apply to you This warranty gives you specific legal rights and you also may have other rights which vary from jurisdiction to jurisdiction This warranty applies only to...

Страница 15: ...een thoroughly tested and inspected to assure proper performance and operation 4 This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is nece...

Страница 16: ...ME SYSTEM may be powered at any NOMINAL input voltage but with the same frequency as the test signal see Table 1 f Before modulation is applied g This test level assumes a minimum distance between the...

Страница 17: ...MA 850 LTE Band 5 Pulse modulationb 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM1800 CDMA1900 GSM1900 DECT LTEBand1 3 4 25 UMTS Pulse modulationb 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 8...

Страница 18: ......

Страница 19: ...Manual del Usuario REF VNT275US Term metro Infrarrojo Corporal Sin Contacto...

Страница 20: ...Term metro Infrarrojo Corporal Sin Contacto es un dispositivo m dico y se puede utilizar repetidamente durante 1 a o Por favor lea el manual antes de utilizar el dispositivo si no entiende completame...

Страница 21: ...secci n Cuidado y Limpieza NOalmaceneesteterm metroenunatemperaturainferiorde 20 Cosuperiora55 C extremaspordebajo de 4 Fom sde131 F oenhumedadexcesiva hastaun93 dehumedadrelativasincondensaci n Perm...

Страница 22: ...stacional Se define como beb s prematuros a los beb s nacidos antes de completar las 37 semanas de gestaci n Se define como beb s peque os para la edad gestacional a los beb s nacimiento a 4 semanas d...

Страница 23: ...IR adquiere energ a infrarroja de la frente seleccionada Teniendo en cuenta la temperatura ambiente el term metro convierte la se al del Sensor IR en un valor digital y lo muestra en la pantalla LCD...

Страница 24: ...delCuerpo Axilar Rango de Medici n Modo cuerpo 34 0 C 42 9 C 93 2 F 109 2 F Modo superficie 0 C 100 C 32 F 212 F Precisi n de Laboratorio Modo cuerpo 0 4 F 0 2 C para el rango de 35 0 C 42 0 C 95 F 10...

Страница 25: ...que para los term metros de mercurio en vidrio y electr nicos el requisito por el Est ndar ASTM E 667 86 y E 1112 86 es 0 1 C 0 2 F Este dispositivo cumple con los siguientes est ndares IEC 60601 1 E...

Страница 26: ...O Retire la pila del instrumento si no es necesaria durante largos per odos de tiempo para evitar da os en el term metro resultante de una pila con fugas 2 Operaci n 2 1 Instalaci n de la pila PRECAUC...

Страница 27: ...gadas luegopresione elbot ndeEncendido Disparador parainiciarlamedici nyleerlosdatos Elterm metro realizar unaautopruebacontodoslossegmentosmostradosdurantedossegundos Nota 1 Despu s de que se hayan v...

Страница 28: ...para ser utilizados por profesionales Se recomienda mantener este term metro en los ajustes originalmente programados en el dispositivo Ajuste de Alerta de Fiebre F2 Esteajustecambiar enelmomentoenque...

Страница 29: ...pantalla parpadea autom ticamente se apaga Cambielapilaoelproductohasido da ado ynecesitareparaci n Advertencia de Pila Baja Cuando la pila se est agotando la pantalla parpadea el s mbolo de advertenc...

Страница 30: ...dado y Almacenamiento La lente es muy delicada Es muy importante proteger la lente de la suciedad y los da os Use un pa o limpio y suave para limpiar la superficie del dispositivo y la pantalla LCD No...

Страница 31: ...ciones o exclusiones no se apliquen en su caso Esta garant a le confiere ciertos derechos legales espec ficos Es posible que usted tambi n tenga otros derechos legales los que var an seg n la jurisdic...

Страница 32: ...robada e inspeccionada a fondo para asegurar el correcto desempe o y operaci n 4 Esta m quina no debe utilizarse adyacente o apilada con otros equipos y que si es necesario un uso adyacente o apilado...

Страница 33: ...TEMA ME pueden alimentarse a cualquier voltaje de entrada NOMINAL pero con la misma frecuencia que la se al de prueba vea la Tabla 1 f Antes de aplicar la modulaci n g Este nivel de prueba supone una...

Страница 34: ...MA 850 Banda LTE 5 Modulaci n porimpulsosb 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM1800 CDMA1900 GSM1900 DECT LTEBanda1 3 4 25 UMTS Modulaci n porimpulsosb 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bl...

Страница 35: ......

Страница 36: ...enofTroy com Visite www VicksThermometers com Por favor aseg rese de especificar el n mero de modelo 2020 Todos los Derechos Reservados Kaz USA Inc una Empresa de Helen of Troy 400 Donald Lynch Blvd S...

Страница 37: ...ount 34 50 Uncoated Revision 7 Release Date Date 12JAN21 Scale 1 1 Rerelease Date Colors Special Instructions Dielines Do not print Cyan 0 Magenta 0 Yellow 0 Black 100 Helen of Troy Creative Services...

Отзывы: