ORL-85/ORL-85P
39
6 R
ETURN
L
OSS
O
PERATION
P
ASS
/F
AIL
MODE
5.
Press the
[
A
]
key.
6.
Tap the
[
Normalizing
]
button.
7.
Tap the
[
OK
]
button.
The normalization will be performed and the display will switch
to measurement mode.
The display shows HIGH db.
NOTE:
The normalization will be automatically performed for all
wavelengths.
Viewing the normalizing settings
I
The instrument is in
Return Loss
mode.
1.
Press the
[
A
]
key.
2.
Tap the
[
More
]
button.
3.
Tap the
[
Info
]
button.
The normalizing value table is displayed.
Pass/Fail mode
Editing the limit in the
λ
-table
I
The instrument is in
Return Loss
mode.
1.
Press the
[
A
]
key.
2.
Tap the
[
More
]
button.
3.
Tap the
[
Edit Limit
]
button.
The Edit Limit-table is displayed.
4.
Tap the desired wavelength value.
5.
To edit the limit of the preselected wavelength, press the
[
A
]
key and tap the
[
Edit Limit
]
button.
6.
Type in the desired value.
7.
Tap the
[
OK
]
button.
8.
Press the
[
F
]
key twice.
Starting the Pass/Fail mode
I
The limit for the
Pass/Fail
mode is set.
1.
Tap the desired wavelength(s).
2.
Press the
[
A
]
key.
3.
Tap the
[
Pass/Fail
]
button.
J
To switch to the
Return Loss
mode, press the
[
A
]
key & tap
the
[
dB
]
button.