[45]
98MANALIDP10AV Rev. A 8/16/11
Error Messages
Table 6. Error Messages
Message
Interpretation
OVR RNG
Normalized calculation result greater than 2% above calibrated span.
a. Overrange input; correct input condition.
b. Bad span calibration; recalibrate span.
c. Bad sensor connection; check electronics module to sensor connection.
d. Defective or damaged sensor; replace sensor.
UND RNG
Normalized calculation result greater than 2% below calibrated zero.
a. Underrange input; correct input condition.
b. Bad zero calibration; recalibrate zero.
c. Bad sensor connection; check electronics module to sensor connection.
d. Defective or damaged sensor; Replace sensor.
FDB ERR
CRC error detected in Factory Database on startup.
a. Incorrect user database; replace sensor.
b. Bad sensor connection; check electronics module to sensor.
c. Defective or damaged sensor; replace sensor.
UDB ERR
CRC error detected in User Database on startup.
a. Incorrect user database; reconfigure/recalibrate transmitter.
b. Bad sensor connection; check electronics module to sensor.
c. Defective or damaged sensor; replace sensor.
BAD IN1
Normalized raw pressure input outside of limits.
a. Extreme overrange or underrange input; Correct input condition.
b. Bad calibration; recalibrate transmitter.
c. Bad sensor connection; check electronics module to sensor.
d. Defective or damaged sensor; replace sensor.
BAD IN3
Normalized raw temperature input outside of limits.
a. Bad sensor connection; check electronics module to sensor.
b. Defective or damaged sensor; replace sensor.
BAD KEY
Invalid keypress detected
a. Pressing External Zero button with
EX ZERO
disabled or transmitter not
Online.
b. Pressing
Enter
when transmitter is Online.
c. Pressing
Next
or
Enter
while
WAIT
is displayed; try again after
WAIT
message
has cleared.
LOLIMIT
Calibration adjustment has reached lower limit.
a. Improper calibration setup; correct setup.
b. Bad D/A converter; replace electronics module.
HILIMIT
Calibration adjustment has reached upper limit.
a. Improper calibration setup; Correct setup.
b. Bad D/A converter; Replace electronics module.
Содержание IDP10-A
Страница 14: ... 14 98MANALIDP10AV Rev A 8 16 11 ...
Страница 16: ... 16 98MANALIDP10AV Rev A 8 16 11 ...
Страница 26: ... 26 98MANALIDP10AV Rev A 8 16 11 ...
Страница 35: ... 35 98MANALIDP10AV Rev A 8 16 11 ...
Страница 37: ... 37 98MANALIDP10AV Rev A 8 16 11 IDP10 A and IDP10 V Calibration ...
Страница 40: ... 40 98MANALIDP10AV Rev A 8 16 11 IDP10 A and IDP10 V Configuration ...