98MANALIDP10AV Rev. A 8/16/11
Commentary on Figure 24
In general, use the
Next
button to select an item and
Enter
button to specify a selection.
EX ZERO: The External Zero feature allows the optional external zero pushbutton to be disabled
for additional security. To configure this feature, go to
EX ZERO
with the
Next
button and
press
Enter
. Use the
Next
button to select
EXZ DIS
or
EXZ ENA
and press
Enter
.
OUT DIR: To configure the Output Direction, go to
OUT DIR
with the
Next
button and press
Enter
.. Use the Next button to select
FORWARD
(4-20 mA for IDP10-A or 1-5 VDC for
IDP10-V) or
REVERSE
(20-4 mA for IDP10-A or 5-1 VDC for IDP10-V) and press
Enter.
.
OUTMODE: To configure the mode of the output, go to
OUTMODE
with the
Next
button
and press
Enter
. Use the
Next
button to select
LINEAR
,
SQ<1CUT
(square root with cutoff
below 1% of calibrated pressure range), or
SQ<4LIN
(square root with dual slope linear below
4% of calibrated pressure range) and press
Enter
.
NOTE
If you wish the output and display to be in square root, it is necessary to first
configure
OUTMODE
as
LINEAR
and follow the Linear Mode path in Figure 24
to establish the pressure LRV and URV. Then go back and configure
OUTMODE
as one of the square root mode selections and follow the Square Root Mode path.
OUTFAIL: The Outfail feature provides high or low output with certain malfunctions. To con-
figure the fail mode output, go to
OUTFAIL
with the
Next
button and press
Enter
. Use the
Next
button to select
FAIL LO
or
FAIL HI
and press
Enter
.
DAMPING: To configure additional damping, go to
DAMPING
with the
Next
button and press
Enter
. Use the
Next
button to select
NO DAMP
,
DAMP 2
,
DAMP 4
, or
DAMP 8
and press
Enter
.
EGU SEL: To configure engineering units for your calibrated range and display, go to
EGU SEL
with the
Next
button and press
Enter
. Depending on how
OUTMODE
is configured, the
remainder of the configuration takes one of two paths.
If
OUTMODE
was configured as
LINEAR
, use the
Next
button to select one of the following
units:
INH2O
,
INHG
,
FTH2O
,
MMH2O
,
MMHG
,
PSI
,
BAR
,
MBAR
,
G/CM2
,
KG/CM2
,
PA
,
KPA
,
MPA
,
TORR
, or
ATM
and press
Enter
. The display advances to
EGU LRV
.
If
OUTMODE
was configured as
SQ<1CUT
or
SQ<4LIN
, you can specify any custom display
unit up to seven characters in length. The display shows
Percent
with the first character flashing.
Use the
Next
button to step through the library of characters (see Table 4) to select the desired
first character, then press
Enter
. Your selection is entered and the second character flashes.
Repeat this procedure until you have created your new unit name. If the unit name has less than
seven characters, use blanks for the remaining spaces. When you have configured the seventh
space, the display advances to
DSP URV
.
EGU LRV: To configure the LRV, press
Enter
at the prompt
EGU LRV
. Use the
Next
button
to toggle between a space or a minus and press
Enter
. Then use the
Next
button to step through
the library of numerical characters to select the desired first digit, and press
Enter
. Your selection
is entered and the second digit flashes. Repeat this procedure until you have entered your last
digit. Then use the
Next
button to move the decimal point to its desired location and press
Enter
.
EGU URV: Similar to EGU LRV immediately above.
Содержание IDP10-A
Страница 14: ... 14 98MANALIDP10AV Rev A 8 16 11 ...
Страница 16: ... 16 98MANALIDP10AV Rev A 8 16 11 ...
Страница 26: ... 26 98MANALIDP10AV Rev A 8 16 11 ...
Страница 35: ... 35 98MANALIDP10AV Rev A 8 16 11 ...
Страница 37: ... 37 98MANALIDP10AV Rev A 8 16 11 IDP10 A and IDP10 V Calibration ...
Страница 40: ... 40 98MANALIDP10AV Rev A 8 16 11 IDP10 A and IDP10 V Configuration ...