Maintenance
3. How to Replace the Electrode Needle
Electrode needle replacement is made with Electrode needle replacement screwdriver G-7DR
(separately sold) only.
Pull off power supply adapter first before the cleaning of Electrode needle.
Replacement Order:
1-2. Remove the Nozzle.
3. Release the screw with Electrode needle replacement screwdriver.
Change it with new needle (G-7H) and fasten firmly with the same screwdriver.
Place the Nozzle back.
4. Mount the Nozzle firmly.
If it is too tightly fastened, it may damage or break itself,
but if it is not well fastened this nozzle may pop out due to its air blow power.
4. Static Erasing Capacity Check
Measure ion balance and charge voltage dampening time using a charge plate monitor.
Reference: IEC61340-5-1-2, EOS/ESD S3.1
5.Replaceable Parts
Electrode Needle
G-7H
Transformer with Robotic Cable
(AC Adapter)
AD24-ITB-E
Silent Nozzle
BB-Z-SN
11