6.1.6 Configuring Detail Threshold Colors
The Details tab can display cell data in color. The threshold levels that
determine the colors are the same as those for graphs, but Details list
color choices are independent of graph colors.
Intertier Tab
Figure 104. Intertier Tab|File Properties
The Intertier tab allows the user to mark cells in a battery as intertier
cells to indicate they are on the boundary of an intertier connection,
manually or automatically
NOTE:
Be certain Number of Cells on the File|Properties|General
tab is correct before marking intertier cells. Marking intertier
configuration affects intercell resistance calculation. Cells
marked as intertier cells are ignored when calculating high,
average, and low intercell resistances for a series of cells
because intertier cells have much higher intercell resistance
than other cells.
By convention, only the cell with the lower cell number is marked. For
example, if Cell 10 is the last cell of one string and connected to Cell 11,
which is the first cell of the next string, only Cell 10 is marked as an
intertier cell. To configure intertier cells, open a battery file, then open
the Intertier tab.
Содержание Alber Cellcorder CRT-400
Страница 1: ...Cellcorder Battery Analysis System User s Guide 590 2085 501A SL 29400 4200 002_Rev6 ...
Страница 17: ...4200 002 xvii Revision 6 Figure 374 CLC 200 Interface 182 Figure 375 Cellcorder CLC 200 Data Capture 183 ...
Страница 18: ......
Страница 47: ...The IMPORTANT Configuration File Editor 4200 002 29 Revision 6 Figure 58 BLF On The Website ...
Страница 167: ...Generating Reports Five Report Types 4200 002 149 Revision 6 Figure 301 Data Set Detail Report Examples ...
Страница 174: ...Generating Reports Five Report Types 4200 002 156 Revision 6 Figure 316 Threshold Report With Graphs Example ...