![VersaLogic VL-12CT96 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/versalogic/vl-12ct96/vl-12ct96_reference-manual_1006747059.webp)
VL-12CT96/7 Analog & Digital I/O Card
7-1
Calibration
The VL-12CT96/7 is calibrated before shipment. However, it may be desirable to recalibrate the card
after installation, and approximately once per year (depending upon the accuracy requirements of the
application).
Analog Input Calibration
Required Equipment
• A voltmeter with resolution and accuracy to ½ LSB of the input range being used.
• A low noise DC voltage source capable of providing a stable reference standard for calibration.
• A small flat blade screwdriver.
Calibration Procedure
• Jumper card for single ended input mode.
• Disconnect cable from connector J1. Use a shorting jumper to short pins 1 and 3 on connector
J1. This applies 0.000 volts (AGND) to channel 0.
• Using a program to continuously read channel 0, adjust the ZERO pot until the reading cen-
ters on 0000H.
• Attach the voltmeter to the voltage source, and adjust (per the table below) to the calibration
voltage appropriate for the input range being used.
• Using a program to continuously read channel 0, adjust the “GAIN” pot until the reading
centers on 07FBH (VL-12CT96) or 7FFBH (VL-12CT97).
————————————Input Range————————————
±5V
±10V
±50mV
±100mV
12 Bit Resolution
+4.987793
+9.975586
+0.049878
+0.099756
16 Bit Resolution
+4.999237
+9.998474
+0.049992
+0.099984
Figure 7-1. Analog Input Calibration Voltages
Calibration
Содержание VL-12CT96
Страница 2: ...ii VL 12CT96 7 Analog Digital I O Card ...
Страница 26: ...2 20 VL 12CT96 7 Analog Digital I O Card ...
Страница 32: ...3 6 VL 12CT96 7 Analog Digital I O Card ...
Страница 62: ...7 4 VL 12CT96 7 Analog Digital I O Card ...
Страница 65: ...VL 12CT96 7 Analog Digital I O Card 8 3 VL 12CT96 7 Parts Placement Reference ...