16
2 Loading or source exchange
Source loading and exchange • VEGASOURCE 31, 35
39155-EN-181126
•
Possibly extension rod M6
•
Remove padlock (1)
•
Loosen locking screw (2) with a flat head screwdriver 1.5 x 10
•
Remove the seal of the capstan screw (3) with the side cutting
pliers and pull out the seal wire
•
Unscrew the capstan screw (3) until the fixing strap (4) can be
folded up
•
Fold up the fixing strap (4)
Fig. 13: VEGASOURCE - version D in extraction position
•
If necessary, screw an extension rod (9) with thread M6 into
the thread of the source holder (5). With this extension you can
increase the distance to the source
•
Unscrew source holder (5) with spanner (size 13) and pull it out of
the source container
•
Only when exchanging the source: Insert the blade of a screw-
driver (4 mm) into the slot (6) on the source holder (5) and press
out the source capsule (7) or unscrew the source capsule with a
screwdriver (2 mm). Let the source capsule (7) fall into a shielded
vessel
•
Insert the new source capsule (7) into the source holder (5) until
it snaps in or insert the spring ringe (8) with the threaded version
and screw the source capsule into the source holder (5) (torque
2 ±0.5 Nm). Use a suitable gripping device or a manipulator for
inserting
•
Insert the source holder (5) completely into the source container
and screw in with a spanner (size 13)
Remove source holder
Insert or exchange
source
Insert source holder