54
VEGASON 51K … 53K
Lin.
curve
Linear
Integr
ation
time
0 s
Adjust
ment
Signal
condit
ioning
Scal
ing
Set-up
with
medium
Min.
adjust
at %
XXX.X
Max.
adjust
at %
XXX.X
Signal
condit
ioning
Scal
ing
0 %
corres
ponds
XXXX
100 %
corres
ponds
XXXX
Deci-
mal
point
888.8
prop.
to
Mass
Unit
Kg
Adjustment with medium
Fill the vessel, for example to 10 % and enter
10 % in the menu "Min. adjust“ with the "+“
and "–“ keys. Then fill the vessel, e.g. to 80 %
or 100 % and enter 80 % or 100 % in the
menu "Max. adjust“ with the "+“ and "–“ keys.
2. Conditioning
Under the menu point "Signal conditioning“
you choose the product distance at 0 % and
at 100 % filling. Then you enter the parameter
and the physical unit as well as the decimal
point.
Enter in the menu window "0 % corresponds“
the value of the 0 %-filling. In the example of
the adjustment with the PC and the adjust-
ment software VVO this would be 45 for 45 l.
• Confirm with "OK“.
With the "—>“ key you change to the 100 %
menu. Enter here the figure of your param-
eter corresponding to a 100 %-filling. In the
example 1200 for 1200 l.
• Confirm with "OK“.
If necessary choose a decimal point. How-
ever note that only max. 4 digits can be dis-
played. In the menu "Prop. to“ you choose
the parameter (mass, volume, distance…)
and in the menu "Unit“ the physical unit (kg, l,
ft
3
, gal, m
3
…).
Linearisation:
A linear dependence between percentage
value of the product distance and the per-
centage value of the filling volume is
preadjusted. With the menu "Lin.curve“ you
can choose between linear, spherical tank
and cylindrical tank. The adjustment of one's
own linearisation curve is only possible with
the PC and the adjustment program VVO.
3. Outputs
Under the menu "Outputs“ you determine if,
for example the current output should be
inverted or which parameter should be pro-
vided by the sensor display.