52
VEGASON 51K … 53K
Max.
Min.
100 % (1,270 m) corresp. to
1200 l
Span (4,58 m)
0 % (5,850 m) corresp. to
45 l
Error codes:
E
Hardware failure, electronics defect
E013
No valid measured value
- Sensor in the feeding phase
- Loss of the useful echo
E017
Adjustment span too small
E036
Sensor program not operating
- Sensor must have a new program-
ming (service)
- Fault signal also appears during
programming
Adjustment steps
On pages 56 and 57 you can find the com-
plete menu plan of the adjustment module
MINICOM.
Set-up the sensor in the numbered se-
quence:
1. Adjustment
2. Conditioning
3. Outputs
4. Operating range
5. Meas. conditions
6. Carry out measurement in a pipe only when
measuring in a standpipe.
7. False echo memory (only required when
meas. errors occur during operation).
8. Indication of the useful and noise level
Short explanations to the set-up steps 1 … 8
follow:
1. Adjustment
-
+
ESC
OK
2
4
Tank 1
m (d)
12.345
-
+
ESC
OK
Tank 1
m (d)
12.345
4 ... 20 mA
Set-up
6.3 Adjustment with the adjustment
module MINICOM
As with the PC, you can also adjust the sen-
sor with the small detachable adjustment
module MINICOM. The adjustment module is
thereby plugged into the sensor or into the
external indicating instrument (optional).
For the adjustment with the adjustment mod-
ule all sensor versions (adjustment options)
as with the PC and the adjustment program
VVO are available. The adjustment with
MINICOM, however, is different.
You carry out all adjustment steps with the 6
keys of the adjustment module. A small dis-
play shows you, apart from the measured
value, a short message on the menu point or
on the figure of a menu adjustment.
The information volume of the small display
however cannot be compared with that of the
adjustment program VVO, but you will soon
get used to it and will be able to carry out
your adjustments quickly and directly with
the small MINICOM.
Содержание VEGASON 51K
Страница 1: ...Level and Pressure Operating Instructions VEGASON 51K 53K...
Страница 63: ...VEGASON 51K 53K 63 Notes...