VEGASON 51K … 53K
Mounting boss on dished tank ends
Open vessels
Reference
plane
£
60 mm
1
/
2
vessel radius
Min. meas.
distance
Open vessles
On open vessels the use of the instruments
on a mounting lever is recommended. Mount
the low-weight sensor to a mounting lever.
Flange on dished vessel ends
Reference plane
1
/
2
vessel radium
Mounting and installation
Mounting on short 1
1
/
2
“ or 2“ socket piece
Reference plane
£
60 mm
You can also mount the sensors with 1
1
/
2
“ or
2“ thread to short socket pieces.
Dished tank ends
On dished tank ends, please do not mount
the instrument in the centre but approx. 1/2
vessel radius from the centre. Dished tank
ends can act as paraboloidal reflectors for
the ultrasonic pulses. If the transducer is
placed in the "focus" of the parabolic tank
end, the transducer receives amplified false
echoes. The transducer should be mounted
outside the "focus". Parabolic amplified ech-
oes are thereby avoided.
£
60 mm
Reference plane