6 Set up with the display and adjustment module
SOLITRAC 31 • Profibus PA
43834-EN-211203
Menu
Menu item
Default value
Setup
Sensor address
126
Measurement loop name
Sensor
Isotope
Cs-137
Application
Summation Secondary
Outputs
Deactivated
AI FB1 Channel
Process value
AI FB1 scaling unit
%
AI FB1 scaling
0 % = 0 %
100 % = 100 %
AI FB1 damping
0 s
Lock adjustment
Released
Display
Language
Selected language
Displayed value
Pulse rate (ct/s)
6.5 Parameter adjustment - Point level detection
The instrument is adapted to the application conditions via the pa-
rameter adjustment. The parameter adjustment is carried out with an
adjustment menu.
Caution:
During the first setup or after an instrument reset the instrument starts
with preset standard values. These value are not suitable for your ap-
plication and must be replaced by real values.
Carry out a setup in the sequence described in the following.
The main menu is divided into five sections with the following func-
tions:
Setup:
Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display:
Settings, for example language, measured value display
Diagnosis:
Information, for example, of device status, peak value,
simulation
Additional adjustments:
Instrument unit, reset, date/time, copying
function
Info:
Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not,
you can change the language in the menu item "
Display - Menu
language
".
Instrument start
Main menu
Procedure