After each treatment session discard all excess medication (risk of contamination, potential
loss of potency of medication).
•
The basic Smart Card supplied with the AKITA JET control unit is not designed for inhalation
of medications, and must be used for device testing and training only.
•
To reduce the risk of contamination and strangulation, keep the AKITA JET Inhalation System
and all medications away from small children.
•
The applied part is the nebuliser or the face mask and the first 10 cm of the tube.
•
The AKITA JET Inhalation System should not be used around flammable substances, or in the
presence of a flammable anaesthetic mixture with air, oxygen or nitrous oxide.
•
Do not use AKITA JET Inhalation System in a ventilator circuit
•
Use the AKITA JET Inhalation System in a clean environment free from dirt, dust, pet hair, etc.
•
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
•
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
•
Do not use the Akita Jet in the bathroom.
•
Do not use without the filter
•
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the AKITA
JET nebuliser, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
CAUTION:
•
Only use the AKITA JET Inhalation System according to the EMC (electromagnetic
compatibility) information provided in the section 11.6.
•
The AKITA JET Inhalation System may be affected by portable or mobile radio frequency
communication equipment. In case of malfunction or error messages, increase distance to
your cellular phone. Detailed information is provided in the section 11.6.
•
Only use the AKITA JET Inhalation System with the approved accessories (see section 0). Non-
approved accessories may have a negative effect on the EMC (electromagnetic compatibility).
•
The AKITA JET Inhalation System is not intended for continuous operation. The normal use is
up to 40 minutes.