Troubleshooting
E
RROR
P
OSSIBLE
C
AUSES AND REMEDY
Nothing appears on display.
The display is not illuminated
and the internal compressor
is not operating.
Ensure compliance with the following points:
•
The power plug is plugged in
•
The power supply is connected to the power cord and the AKITA
JET
•
The device has been switched on with the on/off button
You have switched on the
device but after the
message: “Auto-test on…”
the message “Error: Check
connections” appears.
Ensure compliance with the following points:
•
The original help flow tube is used
•
The tube is correctly connected
•
The tube is not kinked
The AKITA JET performs the
self-test, but does not start
nebulisation even though
you have briefly sucked on
the mouthpiece.
Ensure compliance with the following points:
•
The nebuliser is correctly assembled
•
The help flow tube is connected via help flow adapter to the
nebuliser
•
The help flow tube is not kinked
•
The mouthpiece is correctly connected
At every breath “Inhale
slowly” appears.
Ensure compliance with the following points:
•
You follow the given air flow during inhalation and do not inhale
faster
•
The nebuliser tube is correctly connected
During inhalation the display
shows the message “Error:
Overpressure > KEY“
Ensure compliance with the following points:
•
You have not covered the mouthpiece with your tongue.
•
The tubes are correctly connected and not kinked.
During exhalation the display
shows the message:
“Overpressure error: Do not
exhale into device > KEY“
Do not exhale into the mouthpiece. Take the mouth piece out of your
mouth for exhalation.