Guidance and Manufacturer’s Declaration
Electromagnetic Immunity AC Power Port
The AKITA JET nebuliser is intended for use in the electromagnetic environment specified below. The
customer or the user of the AKITA JET nebuliser should assure that it is used in such an environment
Immunity test
IEC 60601-1-2:2014
Test Condition
Compliance
Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radio-frequency
field
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
3 V
Portable and mobile RF communications
equipment should be used no closer to
any part of the AKITA JET nebuliser,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
𝑑𝑑
=
6
𝐸𝐸
√𝑃𝑃
where
p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer,
E is the filed strength in
V/m and
d is the recommended
separation distance in metres (m).
b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
6 Vrms
ISM/amateur Bands
6V
Fast transient/ burst
IEC 61000-4-4
± 2 kV
for power supply
lines
± 2 kV
Mains power quality should be that of a
typical commercial or hospital
environment.