AKITA
®
JET
36
Radio-frequency
EM fields and
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
80 MHz – 2.7 GHz
80 % AM at 1 kHz
80 MHz – 2.7 GHz Mains power quality should be that of a
typical home healthcare or hospital
environment.
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
27 V/m
380 - 390 MHz
50% PM at 18 Hz
27 V/m
380 - 390 MHz
28 V/m
430 - 470 MHz
50% PM at 18 Hz
28 V/m
430 - 470 MHz
9 V/m
704 - 787 MHz
50% PM at 217 Hz
9 V/m
704 - 787 MHz
28 V/m
800 - 960 MHz
50% PM at 18 Hz
28 V/m
800 - 960 MHz
28 V/m
1700 - 1990 MHz
50% PM at 217 Hz
28 V/m
1700 - 1990 MHz
28 V/m
2400 - 2570 MHz
50% PM at 217 Hz
28 V/m
2400 - 2570 MHz
9 V/m
5100 - 5800 MHz
50% PM at 217 Hz
9 V/m
5100 - 5800 MHz
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.