Use standard dishwashing detergent (for manual dish washing, neutral pH between 6 and 8)
without disinfection properties and without skincare additives (no balsam).
•
Do not use of other cleaning detergents that may cause severe damage to the device or
impose health hazards such as detergents that have disinfectants, detergents with skin care
additives, detergents for dish washing machines, acids (such as vinegar), organic solvents
(such as isopropyl alcohol), or oxidizing agents (such as bleach).
Recommended Procedure:
•
Fill a clean bowl with warm tap water (minimum temperature 35°C). Add standard dish washing
detergent (dosage according to the manufacturer's recommendations).
•
Place the nebuliser components into the bowl. Clean all parts of the nebuliser thoroughly for
5 minutes using a soft brush or soft clean cloth.
•
Then rinse all parts thoroughly under warm running water.
•
To remove excess water you may shake the parts.
•
Drain and place all parts on a new paper towel and let dry completely, or
•
Continue with disinfection (see section 5.3).
5.3
Disinfection
The AKITA JET nebuliser must be disinfected once a day.
Recommended Procedure: Steam disinfector (Baby bottle disinfector)
•
After cleaning the device according to the instructions above, place all components into the
steam disinfector. Make sure that the nebuliser remains disassembled.
•
Operate the bottle disinfector according to the instructions of the manufacturer. Select a
minimum disinfection duration time of 6 minutes.
•
After the disinfection cycle, open the lid of the steam disinfector to allow drying and cooling.
•
Once the components are cool, place them onto a flat dry paper towel.
•
Let the components dry completely (overnight).
Effective cleaning and disinfection were demonstrated by an independent accredited test laboratory
using the dishwashing detergent “Palmolive Original” (Colgate Palmolive GmbH, Hamburg, Germany)
and a steam disinfector (iQ24 Electronic Steam Steriliser, Philips AVENT, Suffolk, England, operated in
mode 1) used according to the recommended procedure described in these Instructions for Use. Use of
alternate disinfector devices or methods is in the responsibility of the user.
Possible Alternative: In boiling water
•
After cleaning the device, fill a saucepan with distilled water.