AKITA
®
JET
18
Finally attach the mouth piece to the nebuliser upper section. Your AKITA JET Inhalation System is now
prepared for the treatment.
CAUTION:
•
Check before every inhalation and during treatment, that the tubes are connected as shown
in the following figure.
Fig. 15: AKITA JET with connected nebuliser.
6.4
Switching on AKITA JET Inhalation System
Before every inhalation, check that the nebuliser is correctly assembled.
Ensure that the device is connected as described in the previous section, that a filter is inserted (see
section 5.6) and the Smart Card is correctly inserted (see section 4.2).
CAUTION:
•
Always ensure when operating the AKITA JET Inhalation System that the ventilation slots at
the back of the control unit remains unobstructed. Never cover the control unit during use.
•
Switch on the AKITA JET Inhalation System with the on/off button.
•
The treatment time for normal use is up to 40 minutes.
Fig. 16: Button for switching on the AKITA JET.
•
Wait until the AKITA JET Inhalation System has completed the self-test.
•
During the self-test, the compressor functionality and the connection of the help flow tube is
tested. During this time, the display shows the following message:
AKITA JET X.X
Auto-test on
Please wait…
•
After the self-test the display will show you the following message: