Appendix
3. System delivery to site:
Yes
No
Did you review and identify the delivery route and method for
equipment in advance?
Is the freight elevator available?
Is the security guard, if any, notified of the installation in advance?
Are two installers, including the helpers, available to move and
unload the equipment?
4. Before installation:
Site check list
Yes
No
Is the room large enough?. At minimum, with CEPH unit
2,200 mm x 2200 mm x 2,600 mm/
87” x 87” x 103. Without CEPH unit, 2,200 mm x 1,400 mm x 2,600
mm/ 87” x 55” x 103”
Is the door entrance wider than 800mm (32”)?
Is a radiation protection plan in place?
Does equipment and PC use same dedicated circuit?
Does the electrical input conditions to installation site
meet the MEIGaN requirements?
Is the local Network IP address of clinic 192.168.33.xx?
Is compressor or air conditioner suction located right next to X-ray
Room?
Is the floor flat and level?
Is the carpet on the floor? If so, remove it
Содержание PaX-i3D Smart
Страница 1: ...PHT 30LFO HARDWARE and SOFTWARE E n g l i s h Installation Manual Version 1 0 0...
Страница 2: ......
Страница 3: ......
Страница 72: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 137: ...10 Technical Specifications...
Страница 139: ...Pax i3D Smart PHT 30LFO Installation Manual 139 10 Technical Specifications Dimension of equipment PANO CT Unit mm...
Страница 140: ...Pax i3D Smart PHT 30LFO Installation Manual 140 10 Technical Specifications PANO CT CEPH Unit mm...
Страница 141: ...Pax i3D Smart PHT 30LFO Installation Manual 141 10 Technical Specifications PANO CT CEPH OneShot type Unit mm...
Страница 145: ...Pax i3D Smart PHT 30LFO Installation Manual 145 Appendix Schematic diagram Components supplied...
Страница 160: ......
Страница 161: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 445 170...