Pax-i3D Smart(PHT-30LFO) Installation Manual
135
9 Acquiring the test image
6.
Select
100 x 70
in the
FOV
drop-down list box.
7.
Subtract -7.1(mm) from FOV 10 x 85 Top position value and type the calculation value in the
Colm-T
box. Ensure the other values(for Bottom/Left/Right) are the same as FOV 100 x 85
values and do not change these values.
If the Adult alignment Top value is -2.840, type -9.940 in the Colm-T
box.
(-2.840 - 7.100= -9.940)
8.
Click
Initial
and wait until the system is initialized.
9.
Click
Capture
when it is enabled.
10.
When Align Sequence window appears, press the X-ray exposure switch
Stay outside of the X-ray shielding room during the exposure.
11.
When
Raw Viewer
window appears, release the X-ray exposure switch.
12
. Select
0009.raw
in the raw file list.
13
. Type
75
in the
T
box
14.
Select
S
in the
M
box.
15.
Make sure that the result meets the CBCT Top collimator alignment standard.
16.
When the Collimator Alignment is completed , click
Close
to exit the Collimator Align program.
Содержание PaX-i3D Smart
Страница 1: ...PHT 30LFO HARDWARE and SOFTWARE E n g l i s h Installation Manual Version 1 0 0...
Страница 2: ......
Страница 3: ......
Страница 72: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 137: ...10 Technical Specifications...
Страница 139: ...Pax i3D Smart PHT 30LFO Installation Manual 139 10 Technical Specifications Dimension of equipment PANO CT Unit mm...
Страница 140: ...Pax i3D Smart PHT 30LFO Installation Manual 140 10 Technical Specifications PANO CT CEPH Unit mm...
Страница 141: ...Pax i3D Smart PHT 30LFO Installation Manual 141 10 Technical Specifications PANO CT CEPH OneShot type Unit mm...
Страница 145: ...Pax i3D Smart PHT 30LFO Installation Manual 145 Appendix Schematic diagram Components supplied...
Страница 160: ......
Страница 161: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 445 170...