9 Acquiring the test image
9.1 Aligning the collimator
9.1.1 PANO collimator Alignment
Adult mode Alignment
1.
Remove the Temple support and Normal bite block from the unit
2.
Run
CollimatorAlign.exe
.
Path: C\VCaptureSW\Util\CollimatorAlign\CollimatorAlign.exe
3.
Click
Align Pano
tab.
4.
Type
50
kVp,
5.0
mA in the KVp/mA filed.
5.
Click
Initial
and wait until the system is initialized.
6.
Click
Capture
when it is enabled.
7.
When Align Sequence window appears, press the X-ray exposure switch
Stay outside of the X-ray shielding room during the exposure.