4. Operation
EzRay Air Wall (VEX-S300W) User Manual
25
EN
G
LISH
3. When tooth types are selected, the exposure angles are automatically set
according to the tooth type. To check the default exposure angles, refer to the
following table.
Tooth Type
Angle of Inclination
Incisor
Maxilla: +40° ~ +50°
Mandible: -22° ~ -28°
Canine
Maxilla: +40° ~ +50°
Mandible: -17° ~ -23°
Molar/Premolar
Maxilla: +25° ~ +35°
Mandible: -2° ~ -8°
Bitewing
+3° ~ +12°
Since the angles of inclination for the maxilla incisor and
canine are the same, the exposure time of the canine is
applied to both the maxilla incisor and canine.
Refer to the following figure to see the angle for molar/premolar.
Содержание EzRay Air Wall
Страница 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 19: ...2 Safety Instructions 10 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 55: ...8 Disposing of the Unit 46 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Страница 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Страница 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Страница 83: ......
Страница 84: ......
Страница 85: ......