Appendix
EzRay Air Wall (VEX-S300W) User Manual
71
EN
G
LISH
Guidance and manufacturer’s declaration - electromagnetic immunity
The VEX-S300W is intended for use in the electromagnetic environment specified
below. The customer or the user of the VEX-S300W should assure that it is used in
such an electromagnetic environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.7 GHz
3 Vrms
150 kHz to
80 MHz
3 V/m 80
MHz to 2.7
GHz
The VEX-S300W must be used only
in a shielded location with a
minimum RF shielding effectiveness
and, for each cable that exits the
shielded location, a minimum RF
filter attenuation of 20 dB from 30
MHz to 230 MHz, 20 dB from 230
MHz to 1 GHz. Field strengths
outside the shielded location from
fixed RF transmitters, as
determined by an electromagnetic
site survey, should be less than 3
V/m. an Interference may occur in
the vicinity of equipment marked
with the following symbol:
NOTE 1) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
NOTE 2) the actual shielding effectiveness and filter attenuation of the shielded
location must be verified to assure that they meet the minimum specification.
Field strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
outside the shielded location in which the VEX-S300W is used exceeds 3V/m, the
VEX-S300W should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as
relocating the VEX-S300W or using a shielded location with a higher RF shielding
effectiveness and filter attenuation.
Содержание EzRay Air Wall
Страница 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 19: ...2 Safety Instructions 10 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 55: ...8 Disposing of the Unit 46 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Страница 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Страница 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Страница 83: ......
Страница 84: ......
Страница 85: ......