4. Operation
EzRay Air Wall (VEX-S300W) User Manual
23
EN
G
LISH
3. After tooth type selection, the patient type should be selected. When the
Adult/Child selection area flickers, turn the jog dial to select the patient type. To
see the Control Panel after selection, refer to the figure below.
After patient type selection
Patient Type
Symbol
Type
Adult
Child
After the tooth type and patient type are selected, the
exposure time is automatically displayed.
4. If you want to change the exposure time, turn the jog dial to adjust the exposure
time from 0.05 to 0.5 s. (increments: 0.01 s)
If you press the jog dial after adjusting the exposure
time in Manual Mode, the exposure time is returned to
the default setting.
To save the exposure time as default in Manual Mode,
press and hold the jog dial for about 3 seconds.
Содержание EzRay Air Wall
Страница 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 19: ...2 Safety Instructions 10 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 55: ...8 Disposing of the Unit 46 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Страница 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Страница 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Страница 83: ......
Страница 84: ......
Страница 85: ......