5. Troubleshooting
40
EzRay Air Wall (VEX-S300W) User Manual
Troubleshooting
Problem
Cause
Solution
Equipment is not
turned on.
The power switch is not
turned on properly.
Turn the equipment power switch
off and turn it back on.
Battery discharged
Recheck after charging the battery
with a charger.
Battery cable is not
connected correctly.
Contact your Service
Representative.
Defective battery
Contact your Service
Representative.
Control Panel is
not turned on.
Defective mainboard
Contact your Service
Representative.
Internal cable
disconnected
Contact your Service
Representative.
No X-ray emission
The generator is cooling. Wait for the cooling time (refer to
‘Duty Cycle’ on
page 70
).
Defective Remote
Exposure Switch
Contact your Service
Representative.
Internal cable
disconnected
Contact your Service
Representative.
Defective generator
Contact your Service
Representative.
Tube lifecycle termination Contact your Service
Representative.
X-ray emission
works, but
exposure is too
light or
completely white.
The equipment has been
positioned incorrectly.
Adjust the position of the
equipment.
Exposure time is too
long.
Decrease the exposure time.
The receptor is facing the
the wrong way.
Reposition the receptor.
X-ray emission
works, but
the exposure is
too dark.
Exposure time is too
short.
Increase the exposure time.
Содержание EzRay Air Wall
Страница 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 19: ...2 Safety Instructions 10 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 55: ...8 Disposing of the Unit 46 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Страница 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Страница 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Страница 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Страница 83: ......
Страница 84: ......
Страница 85: ......