TECHNICAL INFORMATION
54
87-900-884-01 (C)
Connection C
Controller-to-Power Connection
The power supply from the external transformer must
be connected to P3, P8 and P9 connectors of the PCB.
The following figure shows the interconnection details.
1
st
primary
120 Vac 50/60 Hz
2
nd
primary
120 Vac 50/60 Hz
1
st
secondary
18 Vac
0.7 A
2
nd
primary
18 Vac
0.7 A
3
rd
secondary
92 Vac
3 A
4
th
secondary
19 Vac
1 A
Power
310 VA
Mains: 100/120 Vac
Mains: 220/240 Vac
Interconnections Controller - Transformer - Mains
CONTROLLER
Transformer
thermic sensor
CONTROLLER
4
3
2
1
8
7
6
5
Orange
Orange
Jumper
18 Violet
0 Grey
18 Violet
92 Black
0 Black
Note:
A jumper needs to be inserted if the
thermic sensor is not used.
19
0
Red
Red
Orange 120
Brown
100
White
0
Black
120
Blue
0
Transformer
thermic sensor
4
3
2
1
8
7
6
5
Jumper
Orange
Orange
18 Violet
0 Grey
18 Violet
92 Black
0 Black
19
0
Red
Red
Orange 120
Brown
100
White
0
Black
120
Blue
0