TECHNICAL INFORMATION
55
87-900-884-01 (C)
Connection D - Controller-to-Pump Connection
To connect the controller to the pump, an appropriate
interconnection cable must be prepared to connect the
P10 connector of the controller with the pump
connector.
The following figure shows the pump output connector
configuration where pins:
−
A-F = pump temperature sensor
−
B-C-D = 80 Vac 3-phase output to pump motor
stator
−
E = ground.
E
F
A
B
C
D
PHASE R
PHASE S
PHASE T
PUMP TEMP. SENSOR
GND
SCREEN
PUMP TEMP. SENSOR
TURBO-V
PUMP
CONNECTOR
CONTROLLER
PUMP
P10
Controller to pump cable