TECHNICAL INFORMATION

2

87-900-939-01 (D)

DESCRIPTION

The  Turbo-V  300  box  controller  is  a  microprocessor-
controlled,  solid-state,  frequency  converter  with    self-
diagnosis and protection features.

The  controller  drives  the  Turbo-V  300  pump  series  by
controllling the voltage and current respect to the speed
reached by pump.

It incorporates all the facilities required for the operation
of the Turbo-V 300 pump series: pump start/stop, digital
current  and  speed  control,  analog  signals  for  external
indicators.

The power is externally supplied.

All the input/output connections are performed on:

 

9 pin  “D”  type  male  connector  attached  to a  cable

400 mm long for I/O and Electrical DC supply.

 

Pump connection with 400 mm long cable.

 

9 pin “D” type connector for RS232 connection.

TURBO-V 300  75 VDC BOX CONTROLLER
DESCRIPTION

The controller is a solid-state frequency converter which
is  driven  by  a  single  chip  microcomputer  and  is
composed  of  a  PCB  which  includes  all  the  circuitry
necessary for its operation.

The  microcomputer  generates  the  variable  output
voltage  according  to  the  software  and  the  gas  load
condition of the pump.

Moreover, 

it 

manages 

signals 

from 

sensors,

input/output  connection  information,  and  gives  output
for a fully automatic operation.

The  controller  can  be  operated  via  remote  signals
through an RS-232 connection.

The  controller  can  be  operated  in  local  mode  through
suitable  switches  connected  between  the  input  pins  of
the TV300 connector.

CONTROLLER SPECIFICATIONS

Input:
- Voltage

- Current

75 Vdc
with 2 Vpp max ripple

3.0 A max.

Fuse

T 3 A

Output:
- Voltage

- Frequency
- Power

80 Vac nominal ±10%,
3-phase

933 Hz, ±2%

165 W maximum

Compliance to Norms:

-  Radio interferences

-  ESD

-  BURST

-  Radiated RF
   immunity

-  Safety

EN 55011 Class Group 1

EN 61000/4/2

EN 61000/4/4

EN 61000/4/3

EN 61010/1

Installation category

II

Operating temperature

0°C to + 40 °C

Storage temperature

-20°C to + 70°C

Cooling

Internal fan

Weight 

0.5 Kg (1.1 Ibs)

 

!

There can be 75 Vdc voltage referred to ground on the
pump cable or on the serial connector.

WARNING!

Содержание Turbo-V 300 75 Vdc

Страница 1: ...87 900 939 01 D APRIL 2002 INSTRUCTION MANUAL Turbo V 300 75 Vdc Box Controller Model SQ 189 vacuum technologies...

Страница 2: ...Turbo V 300 75 Vdc Box Controller...

Страница 3: ......

Страница 4: ......

Страница 5: ...ump Connector 4 I O Specifications 4 9 pin D Type Connector Pin Assignement 4 Serial Communication Port 5 RS 232 Communication Descriptions 5 Transmission Channel Characteristics 5 Message Structure 5...

Страница 6: ...stand operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maint...

Страница 7: ...and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output voltage according to the software and the gas load condition of th...

Страница 8: ...TECHNICAL INFORMATION 3 87 900 939 01 D CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure Controller outline...

Страница 9: ...ins A F Pump temperature sensor Pin E Ground I O Specifications START STOP START command STOP command Low 0 8 Vdc High 4 to 15 Vdc Analog output Output impedence Minimum load 0 to 10 Vdc proportional...

Страница 10: ...ommunication RS 232 serial port connections Transmission Channel Characteristics levels RS 232 RS 422 baud rate 9600 4800 programmable by a jumper on the board character length 8 bits parity none stop...

Страница 11: ...35 STX ADDR ACK ETX CRC Command SOFT START ON Source PC Destination Inverter 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Inverter Destination PC 02 80 06 03 38 35 STX ADDR ACK E...

Страница 12: ...4 brake 5 normal 6 failure 206 N X ERROR CODE Too high load No connection Short circuit Pump overtemp SoftStart Not Ended Controller overtemp RunUpTime Not Reached Power fail 211 N X PUMP SENSOR TEMPE...

Страница 13: ...9 pin D type connector To deactivate the Low Speed status it is necessary to disconnect pin 9 from pin 15 ground of the same connector The low speed frequency is equal to 622 Hz MAINTENANCE Replacemen...

Страница 14: ...TECHNICAL INFORMATION 9 87 900 939 01 D PCB JUMPERS...

Страница 15: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 16: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 17: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Отзывы: