TECHNICAL INFORMATION

8

87-900-939-01 (D)

OPERATION

Make  all  vacuum  manifold  and  electrical  connections
and  refer  to  Turbo-V  pump  instruction  manual  prior  to
operating the Turbo-V controller.

 

!

To  avoid  injury  to  personnel  and  damage  to  the
equipment, if the pump is laying on a table make sure it
is steady.
Never operate the Turbo-V pump if the pump inlet is not
connected to the system or blanked off.

The  controller  operates  completely  automatically  after
the remote start command is given.

Switching on/off the Pump

To switch on the pump it is necessary to short circuit pin
1 and pins 5-6 (ground) of the 9 pin “D” type connector.

To  switch  off  the  pump  it  is  necessary  to  remove  the
short circuit between pins 1 and 5-6.

Analog output:

0  to  10  Vdc  proportional  to  speed  (0
to 10 V 

 0 to 100% speed).

This  output  is  active  also  during  the
pump “slow down” phase after a Stop
command.

Low Speed Activation/Deactivation

To  activate  the  Low  Speed  status  it  is  necessary  to
connect pin 9 of the 9-pin connector to pin 5-6 (ground)
of the 9-pin “D” type connector.

To  deactivate  the  Low  Speed  status  it  is  necessary  to
disconnect  pin  9  from  pin  15  (ground)  of  the  same
connector.

The low speed frequency is equal to 622 Hz.

MAINTENANCE

Replacement  controllers  are  available  on  an  advance
exchange  basis  through  Varian.  If  necessary,
information  is  provided  to  aid  the  operator  in
determining  malfunctions  and  corrective  steps  to  be
taken.

 

!

In  order  Voltages  developed  in  the  unit  are  dangerous
and  may  be  fatal.  Service  must  be  performed  by
authorized personnel only.

Error Messages

For  a  certain  type  of  failure,  the  controller  will  self-
diagnose  the  error  and  the  following  messages  will  be
displayed.

The controller signals the error occurred by means of a
diagnostic LED located on the box (FAULT), and on the
RS 232 port.

The LED blinks in a coded mode:
it flashes a number of time equal to the error
code (see the following table) and then stays off, and so
on.

“Status” LED (on the box)

OFF

in STOP

Blinking

in STARTING

ON

in NORMAL

Error Code Table

LED

BLINKING

NUMBER

DESCRIPTION

0

No error

1

Output overcurrent

2

Not connected pump

3

Pump overtemperature

4

Controller overtemperature

5

Run-up overtime

6

Soft start overtime

7

Too High Load

8

Power Failure

 

WARNING!

 

WARNING!

Содержание Turbo-V 300 75 Vdc

Страница 1: ...87 900 939 01 D APRIL 2002 INSTRUCTION MANUAL Turbo V 300 75 Vdc Box Controller Model SQ 189 vacuum technologies...

Страница 2: ...Turbo V 300 75 Vdc Box Controller...

Страница 3: ......

Страница 4: ......

Страница 5: ...ump Connector 4 I O Specifications 4 9 pin D Type Connector Pin Assignement 4 Serial Communication Port 5 RS 232 Communication Descriptions 5 Transmission Channel Characteristics 5 Message Structure 5...

Страница 6: ...stand operation maintenance manuals and any additional information provided by Varian All warnings and cautions should be read carefully and strictly observed Address any safety operation and or maint...

Страница 7: ...and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output voltage according to the software and the gas load condition of th...

Страница 8: ...TECHNICAL INFORMATION 3 87 900 939 01 D CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure Controller outline...

Страница 9: ...ins A F Pump temperature sensor Pin E Ground I O Specifications START STOP START command STOP command Low 0 8 Vdc High 4 to 15 Vdc Analog output Output impedence Minimum load 0 to 10 Vdc proportional...

Страница 10: ...ommunication RS 232 serial port connections Transmission Channel Characteristics levels RS 232 RS 422 baud rate 9600 4800 programmable by a jumper on the board character length 8 bits parity none stop...

Страница 11: ...35 STX ADDR ACK ETX CRC Command SOFT START ON Source PC Destination Inverter 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Inverter Destination PC 02 80 06 03 38 35 STX ADDR ACK E...

Страница 12: ...4 brake 5 normal 6 failure 206 N X ERROR CODE Too high load No connection Short circuit Pump overtemp SoftStart Not Ended Controller overtemp RunUpTime Not Reached Power fail 211 N X PUMP SENSOR TEMPE...

Страница 13: ...9 pin D type connector To deactivate the Low Speed status it is necessary to disconnect pin 9 from pin 15 ground of the same connector The low speed frequency is equal to 622 Hz MAINTENANCE Replacemen...

Страница 14: ...TECHNICAL INFORMATION 9 87 900 939 01 D PCB JUMPERS...

Страница 15: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 16: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 17: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Отзывы: