background image

3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use 

Page  7

Section 1   Indications, Warnings and Cautions

• 

If using the Oxygen Assist Module during full body irradiation, keep the SpO

2

 sensor out of the 

radiation field. If the sensor is exposed to the radiation, the reading might be inaccurate or the 

device might read zero for the duration of the active irradiation period.

• 

Variation in measurements may be profound and may be affected by sampling technique as 

well as the patient's physiological conditions. Any results exhibiting inconsistency with the 

patient’s clinical status should be repeated and/or supplemented with additional test data. 

Blood samples should be analyzed by laboratory instruments prior to clinical decision making 

to completely understand the patient’s condition.

• 

Do not submerge the Oxygen Assist Module or any associated accessories in any cleaning 

solution or attempt to sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any 

other method. This will seriously damage the device.

• 

Electrical Shock Hazard: Carry out periodic tests to verify that leakage currents of patient-

applied circuits and the system are within acceptable limits as specified by the applicable 

safety standards. The summation of leakage currents must be checked and in compliance 

with IEC 60601-1 and UL60601-1. The system leakage current must be checked when 

connecting external equipment to the system. When an event such as a component drop of 

approximately 1 meter or greater or a spillage of blood or other liquids occurs, retest before 

further use. Injury to personnel could occur.

• 

Disposal of product - Comply with local laws in the disposal of the device and/or its 

accessories.

• 

To minimize radio interference, other electrical equipment that emits radio frequency 

transmissions should not be in close proximity to the Oxygen Assist Module.

• 

Replace the SpO

2

 cable or sensor when a replace sensor or when a low SIQ message is 

consistently displayed while monitoring consecutive patients after completing troubleshooting 

steps listed in this manual.

Notes:

• 

A functional tester cannot be used to assess the accuracy of the SpO

2

 measurement on the 

Oxygen Assist Module.

• 

High-intensity extreme lights (such as pulsating strobe lights) directed on the SpO

2

 sensor, 

may not allow the Oxygen Assist Module to obtain an accurate reading.

• 

Do not loop the SpO

2

 patient cabling into a tight coil or wrap around the device, as this can 

damage the patient cabling.

• 

Additional information specific to the Masimo sensors compatible with the Oxygen Assist 

Module, including information about parameter/measurement performance during motion and 

low perfusion, may be found in the sensor's directions for use (DFU).

• 

SpO

2

 cables and sensors are provided with X-Cal™ technology to minimize the risk of 

inaccurate readings and unanticipated loss of patient monitoring. Refer to the Cable or Sensor 

DFU for the specified duration of the patient monitoring time.

This device is covered under one or more patents as set forth at: www.masimo.com/patents.htm
NO IMPLIED LICENSE: Possession or purchase of this device does not convey any express or 

implied license to use the device with unauthorized sensors or cables which would, alone or in 

combination with this device, fall within the scope of one or more of the patents relating to this 

device.

Содержание Oxygen Assist Module

Страница 1: ...Oxygen Assist Module for use with Vapotherm Precision Flow Instructions for Use...

Страница 2: ...ection 7 Patient Admission 12 Section 8 Alarm Setup 13 Section 9 SpO2 Setup 13 Section 10 O2 Setup 16 Section 11 System Shutdown 17 Section 12 Graphical Trend Display 17 Section 13 Logging of Variable...

Страница 3: ...osed as an unsorted municipal waste and must be collected separately Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment De...

Страница 4: ...y Therapy Equipment and Guidelines for Prevention of Nosocomial Pneumonia For more information see the Precision Flow Instructions for Use General Warnings Federal Law U S restricts the use of this de...

Страница 5: ...rous oxide To ensure safety avoid stacking multiple devices or placing anything on the device during operation To protect against injury follow the directions below Avoid placing the device on surface...

Страница 6: ...lly calibrated in healthy adult volunteers with normal levels of carboxyhemoglobin COHb and methemoglobin MetHb Do not adjust repair open disassemble or modify the Oxygen Assist Module or accessories...

Страница 7: ...her use Injury to personnel could occur Disposal of product Comply with local laws in the disposal of the device and or its accessories To minimize radio interference other electrical equipment that e...

Страница 8: ...sis of real time measurements and trending to choose the appropriate O2 delivery The use of smart averaging and hysteresis algorithms by the Masimo SET SpO2 monitor protects the system from artifacts...

Страница 9: ...ssist Module and provides 12VDC to the Oxygen Assist Module Contact Vapotherm if you believe your power adapter to not be functional or you need a replacement Only use power adapter as provided by Vap...

Страница 10: ...w s interface port 6 Plug the Oxygen Assist Module Power Supply into the Oxygen Assist Module and into an electrical outlet marked Hospital Grade or Hospital Only NOTE Ensure proper orientation on Pow...

Страница 11: ...oxygen Using Menus To call up a menu press the desired menu tab Alternately parameter boxes will redirect to the associated menu Notes Unless otherwise stated the change to settings takes effect when...

Страница 12: ...tely place the Oxygen Assist Module in Manual O2 Mode A manual change in O2 on the Precision Flow with 120 90 60 30 15 5 seconds selected will suppress Auto Mode O2 titration for that period of time b...

Страница 13: ...rm is generated stating No Sensor Connected Alarm messages cycle in the message area For a list of alarm messages and required action please see the section Alarm and Status Messages For alarm indicat...

Страница 14: ...ors may cause incorrect SpO2 readings Before using the sensor carefully read the sensor documentation including all warnings cautions and instructions The accuracy of pulse oximetry readings can be ad...

Страница 15: ...es Signal IQ values of less than 0 3 The SpO2 display range is from 1 100 A value of indicates that no measurement is available The pulse rate display range is from 1 240 bpm A value of indicates that...

Страница 16: ...trol knob Note See description of Fallback O2 Functionality on the following page for more information 3 Insert Oxygen Assist Module Time Card and Press Start Case When the patient is stable press the...

Страница 17: ...s If the Oxygen Assist Module has exited the Auto mode check the patient sensor placement connections and settings When clinically indicated place the Oxygen Assist Module back into Auto mode in the P...

Страница 18: ...ng and disinfecting instructions for the SpO2 sensor and patient cable see the instructions provided with the sensor Section 14 Cleaning and Disinfecting Section 17 Referenced Documentation Precision...

Страница 19: ...Clinical patient condition Check patient Sensor not connected Check sensor and cable connections Sensor off patient Check sensor application Sensor or cables damaged Check sensor and cables Replace i...

Страница 20: ...yperbaric cond Storage and Shipping Ambient Temperature 10 50 C Ambient Relative Humidity 20 90 RH Contact with Patient Indirectly via SpO2 sensor and Precision Flow O2 cannula Liquid Ingress Protecti...

Страница 21: ...trical Equipment Part 1 6 General requirements for basic safety and essential performance Collateral Standard Usability IEC 60601 1 8 2006 Second Edition Am 1 2012 for use in conjunction with IEC 6060...

Страница 22: ...female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non repetitive motion bet...

Страница 23: ...Case Check SpO2 Defective Cable Replace cable enter MANUAL Mode or End Case Check SpO2 Defective Sensor Replace sensor enter MANUAL Mode or End Case Check SpO2 Incompatible Adhesive Sensor Replace se...

Страница 24: ...Demo Mode Reconnect sensor with good signal Check SpO2 Interference Detected Reconnect sensor with good signal Check SpO2 Low Perfusion Index Reconnect sensor with good signal Check SpO2 Low Signal I...

Страница 25: ...target and sets the FiO2 delivery rate accordingly The algorithm makes near immediate adjustments within 10 seconds to changes in SpO2 readings learns from past changes and anticipates short term cha...

Страница 26: ...s declaration electromagnetic emissions The Oxygen Assist Module is intended for use in the electromagnetic environment specified below The customer or the user of the Oxygen Assist Module should ass...

Страница 27: ...y should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 0 5 1 and 2 kV Line s to Ground 1 kV line s to Line s 0 5 1 and 2 kV 1 kV Mains power quality should be that of a t...

Страница 28: ...paration distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interferenc...

Страница 29: ...ed maximum output power of transmitter W Separation distance according to frequency of transmitter m 50 kHz to 80 MHz outside ISM bands d 1 2 P 150 kHz to 80 MHz in ISM bands d 1 2 P 800 MHz to 2 5 GH...

Страница 30: ...Page 30 3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use This Page Intentionally Blank...

Страница 31: ...3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use Page 31 This Page Intentionally Blank...

Страница 32: ...for Use Vapotherm Inc 100 Domain Drive Exeter NH 03833 USA Phone 603 658 0011 Fax 603 658 0181 RMS UK Limited 28 Trinity Road Nailsea North Somerset BS48 4NU United Kingdom Phone 44 1275 85 88 91 Fax...

Отзывы: