Page 24
3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use
Section 20 Alarm and Advisory Messages
Silent Advisory Message
Action and Troubleshooting
%O
2
at user set %O
2
limit*
Increase %O
2
Alarm Limit, or enter MANUAL mode
Battery Backup In Use
Oxygen Assist Module is in Battery Mode. When external
power is lost, the module will automatically switch to Battery
Mode.
Battery Low
Plug the Oxygen Assist Module into external power
Check Precision Flow
Clear Precision Flow alarm and ensure it is in Run Mode with
both Air and O
2
sources present
Check SpO
2
- Adhesive Sensor Near Expiration
Replace SpO
2
Adhesive Sensor
Check SpO
2
- Cable Expired*
Replace Patient Cable
Check SpO
2
- Cable Near Expiration
Replace SpO
2
Patient Cable
Check SpO
2
- Check Sensor Connection*
Reconnect sensor with good signal
Check SpO
2
- Demo Mode*
Reconnect sensor with good signal
Check SpO
2
- Interference Detected*
Reconnect sensor with good signal
Check SpO
2
- Low Perfusion Index*
Reconnect sensor with good signal
Check SpO
2
- Low Signal IQ*
Reconnect sensor with good signal
Check SpO
2
- Pulse Search*
Reconnect sensor with good signal
Check SpO
2
- Sensor Fault*
Reconnect sensor with good signal
Check SpO
2
- Sensor Initializing*
Reconnect sensor with good signal
Check SpO
2
- Sensor Near Expiration
Replace SpO
2
Sensor
Check SpO
2
- Sensor Off Patient*
Reconnect sensor with good signal
Check SpO
2
- SpO
2
Only Mode*
Reconnect sensor with good signal
Local Storage is Low
Delete Previous Case Data Files from Local Storage
Local Storage is Full
Delete Previous Case Data Files from Local Storage
SpO
2
Target > 95%
Change SpO
2
Target to 95% or Below
SpO
2
Target < 85%
Change SpO
2
Target to 85% or Above
Time Near Expiration
Insert new Time Card before time expires or End Case
* These Silent Advisory Messages will become Medium Priority Alarms if not corrected after 2 minutes.