3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use
Section 10 %O
2
Setup
Goal and purpose:
The automatic %O
2
control system uses the patient’s measured SpO
2
values to
control the delivery of %O
2
based on patient need. When the automatic %O
2
control mode is enabled,
the system tries to maintain the patient’s SpO
2
level at or near a preconfigured SpO
2
target (see also
Further Reading).
1.
Select the
Patient
menu to set the %O
2
Mode.
2.
Press the Backup %O
2
parameter box on the touch screen, and set the patient’s Backup %O
2
using the settings control knob.
Note:
See description of Fallback %O
2
Functionality on the following page for more information.
3.
Insert Oxygen Assist Module Time Card and Press Start Case. When the patient is stable, press
the O
2
Mode AUTO button to put the device into Auto Mode (default is Manual).
Note: To enable Auto %O
2
Mode, there must be a time card inserted and a patient case started
on the Oxygen Assist Module.
4.
The Home screen will now display AUTO mode and the Oxygen Assist Module will begin auto
%O
2
titration based on the patient’s SpO
2
level and the set %SpO
2
Target. %O
2
delivered by the
Precision Flow can range from 21-100%, unless a %O
2
Limit is set. Delivered %O
2
can be seen on
the Precision Flow display, or on the Oxygen Assist Module Home Screen graphical trend display.