Section 8 Alarm Setup
The Oxygen Assist Module announces 1 level of alarm:
•
Medium Priority Alarm: Flashing amber; burst of 3 beeps.
The Oxygen Assist Module displays silent advisory messages in the message area.
Notes:
Before starting a new case, with the patient sensor disconnected, confirm that the alarm system is
functional by observing that an alarm is generated stating "No Sensor Connected".
Alarm messages cycle in the message area.
For a list of alarm messages and required action, please see the section Alarm and Status
Messages.
For alarm indications of the Precision Flow, please see the Precision Flow Instructions for Use.
Alarm Silence/Suspend
•
Press the Alarm Silence fixed key to mute all alarms for 2 minutes.
Note:
If alarms are muted, a crossed bell icon appears in the upper right-hand
corner of the screen.
Warnings:
If alarms are silenced for 2 minutes, a subsequent alarm of the same alarm priority will not be
announced during the 2-minute alarm silence period.