27
4.2.4 TCC – Temperature Compensator Calibration
This procedure is used to match a 4176 with its Omni Compensator.
The Temperature Compensator Calibration must be done in a
temperature controlled area.
When the readings stabilize, the instrument will automatically skip to the next screen.
Using the numeric keys input a value for the reference temperature followed by the
E
NTER
key. For corrections use the
C
LEAR
key.
The new values will be stored in memory.