Chapter 7 ___________________________________________________ Retrieving Data Messages
VAISALA______________________________________________________________________ 111
Command format:
aRx!
The response:
a+<data fields><cr><lf>
NOTE
For using Continuous measurement commands for all WXT530 series
parameters (wind, PTU, precipitation, and supervisor) the select the
respective protocol (
aXU,M=R
).
The
M=S
selection requires use of
aM
,
aMC
,
aC
,
aCC
+
aD
commands, only the precipitation data can be retrieved continuously
(using
aR3
command).
where
a
=
Device address
R
=
Start continuous measurement command:
x
=
The desired sensor to make the measurement:
1 = Wind
2 = Temperature, humidity, pressure
3 = Precipitation
5 = Supervisor
If x is left out, the query refers to the combined data
message used for requesting data from several
sensors with just one command.
!
=
Command terminator
where
a
=
Device address
<data
fields>
=
The measured parameters in selected units, separated
with '+' marks (or '-' marks in case of negative
parameter values). The maximum number of
parameters to be measured with one request is 15.
<cr><lf>
=
Response terminator