MI instruction manual
Document/Revision No.
IM-EN-MI
/2.00
Effective: September 2, 2019
3.2 Refractometer cable connections
The refractometers are connected to the MI with a M12 male-female 8 pins A-coded
cable PR-8430, maximum length 90 m. The refractometer connectors are located un-
derneath the MI enclosure as a group of four. Refractometer connections can be made
while the MI is powered. A refractometer can be connected to any of the four connec-
tors.
3.3 Ethernet cable connection
The leftmost connector underneath the MI is for ethernet output. It is for a M12 male
4-pins D-coded cable. The cable can be connected while the MI is powered.
If you wish to connect the MI with e.g. a laptop, you can purchase Interconnecting
cable PR-8330 with a M12-4 pin connector in one end and a RJ-45 connector in the
other end.
3.4 Module cards
For the installation of the module cards, see Chapter 6 “Module cards”.
Содержание MI
Страница 1: ...IM EN MI v 2 00 Instruction Manual Multichannel User Interface MI ...
Страница 4: ...Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 8: ...MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 11: ...2 Mounting 3 Copyright Vaisala 2019 All rights reserved Figure 2 2 Enclosure dimensions ...
Страница 83: ...8 Specifications 75 Copyright Vaisala 2019 All rights reserved 8 2 Parts ...
Страница 84: ...76 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 86: ...78 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 88: ...80 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 90: ...82 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Страница 92: ...www vaisala com ...