User's Guide _______________________________________________________________________
76 ___________________________________________________________________ M211399EN-G
1209-028
Figure 37
DIP Switches in Factory Reset Position (HMW95)
4.
Reconnect the transmitter cover to the mounting base so it powers
up. Check the screen after power-up: when the DIP switches are in
factory reset position, you will see a notification text.
5.
Disconnect the transmitter cover again.
6.
Set the DIP switches to the positions they were before.
7.
Reconnect the transmitter cover to the mounting base. Check the
startup screens to verify the configuration.
HMW95
1
ON
2 3 4 5 6 7 8
Modbus
BACnet
Parity Even
A B C
Parity None
Baud
Rate
Metric
Non-Metric