User's Guide _______________________________________________________________________
50 ___________________________________________________________________ M211399EN-G
Example (show currently displayed parameters):
>
pass 9000
>
dsel
Quant 1 : T
Quant 2 : RH
Quant 3 : Disabled
Example (change display to only show RH):
>
dsel RH
OK
Example (change display to show same parameters as are assigned to
analog output channels):
>
dsel out1 out2
OK
Serial Line Output Commands
Start Measurement Output
Use the
R
command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the
FORM
command.
R
<cr>
Example (measurement message in default format):
>
r
RH = 21.71 %RH T = 23.13 'C
RH = 21.72 %RH T = 23.12 'C
RH = 21.77 %RH T = 23.12 'C
RH = 21.77 %RH T = 23.12 'C
...
Outputting the results continues in intervals issued with the command
INTV
. You can stop the output with the
S
command. Since the interface
is half-duplex, you must enter the commands when the transmitter is not
outputting.
Stop Measurement Output
You can stop the measurement output with the
S
command:
S
<cr>