User's Guide ______________________________________________________________________
86 __________________________________________________________________ M010056EN-G
Accuracy of Mixing Ratio g/kg
(Ambient pressure 1013 mbar)
Accuracy of Wet Bulb Temperature °C
Relative humidity
Temp. 10
20
30
40
50
60
70
80
90
100
-40
0.003
0.003
0.003
0.003
0.003
0.004
0.004
0.004
—
—
-20
0.017
0.018
0.019
0.021
0.022
0.023
0.025
0.026
—
—
0
0.08
0.09
0.09
0.1
0.1
0.11
0.11
0.12
0.13
0.13
20
0.31
0.33
0.35
0.37
0.39
0.41
0.43
0.45
0.47
0.49
40
0.97
1.03
1.1
1.17
1.24
1.31
1.38
1.46
1.54
1.62
60
2.68
2.91
3.16
3.43
3.72
4.04
4.38
4.75
5.15
5.58
80
6.73
7.73
8.92
10.34
12.05
14.14
16.71
19.92
24.01
29.29
100
16.26
21.34
28.89
40.75
60.86
98.85
183.66 438.56 —
—
120
40.83
74.66
172.36 —
—
—
—
—
—
—
Relative humidity
Temp. 10
20
30
40
50
60
70
80
90
100
-40
0.2
0.2
0.2
0.2
0.2
0.2
0.2
0.2
—
—
-20
0.21
0.21
0.22
0.22
0.22
0.22
0.23
0.23
—
—
0
0.27
0.28
0.28
0.29
0.29
0.29
0.3
0.3
0.31
0.31
20
0.45
0.45
0.45
0.44
0.44
0.44
0.43
0.43
0.42
0.42
40
0.84
0.77
0.72
0.67
0.64
0.61
0.58
0.56
0.54
0.52
60
1.45
1.2
1.03
0.91
0.83
0.76
0.71
0.67
0.63
0.6
80
2.23
1.64
1.32
1.13
0.99
0.89
0.82
0.76
0.72
0.68
100
3.06
2.04
1.58
1.31
1.14
1.01
0.92
0.85
0.8
0.75
120
3.85
2.4
1.81
1.48
1.28
1.13
1.03
0.95
0.88
0.83
140
4.57
2.73
2.03
1.65
1.41
1.25
1.13
1.04
0.97
0.91
160
5.25
3.06
2.25
1.82
1.55
1.37
1.24
1.13
1.05
0.99