Chapter 6 __________________________________________________ Calibration and Adjustment
VAISALA _______________________________________________________________________ 69
3.
Let the sensor stabilize; use a multimeter or the transmitter display
to monitor the stabilization.
4.
Press buttons
Up
and
Dn
to adjust the display or multimeter
reading to correspond to the reference value (when using a
multimeter calculate the current value corresponding to the
reference humidity by using the equations presented in
Correspondence of Current Values and Output Quantities on page
64
). Conclude the one-point adjustment by pressing button
E
twice.
ACAL Analog Output Calibration
Syntax:
ACAL
<cr>
Connect HMT360 to a multimeter. Issue the
ACAL
command:
1.
Disconnect the serial cable from the transmitter while reading the
multimeter value for the Ch 1 (I1).
2.
Reconnect the serial cable. Type the multimeter reading and press
Enter.
3.
Type the higher current multimeter reading and press Enter.
Example:
>acal<cr>
Ch1 I1 ( mA ) ? 4.846<cr>
Ch1 I2 ( mA ) ? 19.987<cr>
>