User's Guide ______________________________________________________________________
54 __________________________________________________________________ M010056EN-G
Example:
>ct<cr>
T : 0.811 1. ref ? 0.5<cr>
Press any key when ready ...<cr>
T : 99.122 2. ref ? 99.5<cr>
OK
>
The OK indicates that the adjustment was succesful. When performing
one-point adjustment press only Enter for the second reference.
Output Commands
ITEST Testing Analog Outputs
Syntax:
ITEST
aa.aaa
bb.bbb
<cr>
This command outputs the current value of each channel and the
corresponding control signal of the digital-to-analog converter.
Example:
>itest 8 12<cr>
8.00000 403 12.00000 7DF
>itest<cr>
7.00150 30A 11.35429 73E
>
The set current values remain valid until you issue the command
ITEST
without readings or reset the transmitter. With this command the desired
outputs of the transmitter are shown.
NOTE
Before giving command
ITEST
, reset the transmitter with command
RESET
RESET Transmitter Reset on page 57
where
aa.aaa =
Current value to be set for Ch 1 (mA)
bb.bbb =
Current value to be set for Ch 2 (mA)