User's Guide ______________________________________________________________________
80 __________________________________________________________________ M010056EN-G
Calculated Variables (Typical
Ranges)
With HMP361 Probe
With HMP363, HMP364, HMP365, HMP367 and
HMP368 Probes
Outputs
Table 10
HMP361 Calculated Variables Specifications
Property
Description / Value
Dewpoint temperature
-40 ... +60 °C
Mixing ratio
0 ... 160 g/kg d.a.
Absolute humidity
0 ... 160 g/m³
Wet bulb temperature
0 ... +60 °C
Table 11
HMP363, HMP364, HMP365, HMP367& HMP368
Probes Calculated Variables Specifications
Property
Description / Value
Dewpoint temperature
-40 ... +100 °C
Mixing ratio
0 ... 500 g/kg d.a.
Absolute humidity
0 ... 600 g/m³
Wet bulb temperature
0 ... +100 °C
Table 12
Output Specifications
Property
Description / Value
Two analog outputs (one standard, one
optional)
two wire 4 ... 20 mA
Typical accuracy of analog outputs at
+20 °C
± 0.05 %/°C full scale
Typical temperature dependence of analog
outputs
± 0.005 %/°C full scale
RS232C serial output
for service use only
in safe area
connector type RJ45
WARNING
Output connections are made via safety barriers.