Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 55
SEND Outputting Measurement Values
Syntax:
SEND
<cr>
This command outputs the measured values in one point.
R Activating Continuous Output
Syntax:
R
<cr>
With the command
R
the transmitter outputs measured values
continuously.
S Stopping Continuous Output
Syntax:
S
<cr>
The continuous outputting is stopped with the command
S
. If outputting
is active, this command is not echoed.
INTV Setting Output Interval
Syntax:
INTV
n
xxx
<cr>
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is active.
Example, the output interval is set to 10 minutes:
>intv 10 min<cr>
Output interval: 10 MIN
>
NOTE
When outputting low currents from Ch 1, remember to remove
RS232C-cable while reading the current output, because of the
increased current consumption for using RS port.
where
n
=
1 ... 255
xxx =
S, MIN or H