Replacing diaphragms and valves

All bearings are encapsulated and are filled with long-life lubricant. Under normal oper-
ating conditions, the pump is maintenance free. The valves and the diaphragms as well
as the motor capacitors are wear parts. If the rated ultimate vacuum is no longer
achieved or in case of increased noise level, the pump interior, the diaphragms and the
valves must be cleaned and the diaphragms and valves must be checked for cracks or
other damage.
Depending on individual cases it may be efficient to check and clean the pump heads
on a regular basis. In case of normal wear the lifetime of the diaphragms and valves is
> 10000 operating hours.

Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are pumped in significant amount.

If the pump is exposed to corrosive gases or vapour or in case of deposits, mainte-
nance should be carried out frequently.

Regular maintenance will improve the lifetime of the pump and also protect both
man and environment.

Before starting maintenance

 vent the system, isolate the pump and other compo-

nents from the vacuum system and the electrical supply. Drain condensate if applica-
ble, avoid the release of pollutants. Separate pump from cooling water circuit. Allow
sufficient cooling of the pump. 

Wait two minutes

 after isolating the equipment from

mains to allow the capacitors to discharge.

 Ensure that the pump cannot be oper-

ated accidentally. Never operate the pump if covers or other parts of the pump are
disassembled. Never operate a defective or damaged pump.

Ensure that the maintenance technician is familiar with the safety procedures which
relate to the products processed by the pumping system.

The pump might be contaminated with the process chemicals that have been pumped
during operation. Ensure that the pump is decontaminated before maintenance and
take adequate precautions to protect people from the effects of dangerous substances
if contamination has occurred.

Wear appropriate safety-clothing when you come in contact with contaminated com-
ponents.

Tools required (metric):

-

Phillips screw driver size 2

-

Open-ended wrench w/f 10/14/17

-

Hex key size 5/7

-

Face wrench with torque indicator

Set of seals 

(2 diaphragms and 4 valves) for MZ 2C 

(PC 500)

............ ........................................

696814

Set of seals

 (4 diaphragms and 8 valves) for MD 4C

 (PC 600)

 ............ .......................................

696815

Face wrench

 with torque indicator ...............................................................................................

637580

Please read section ”Replacing diaphragms and valves” completely before start-
ing maintenance.

Partially the pictures show pumps in other versions. This doesn´t influence replacing diaphragms and
valves of the pump.

Содержание PC 500 LAN

Страница 1: ...changed It is strictly the users responsibility to check carefully the validity of this document with respect to his product Manual no 999084 05 10 2009 page 1 of 46 Technology for Vacuum Systems Inst...

Страница 2: ...pump VACUUBRAND diaphragm pumps are the result of many years of experience in construction and practical operation of these pumps combined with the latest developments in material and manufacturing t...

Страница 3: ...ents Safety information 4 Technical data 8 Use and operation 12 Description 16 Notes on operation 18 General view of factory set modes 19 Menu structure of controller 20 Working with the controller 22...

Страница 4: ...ical shock The supply cable may be fitted with a moulded European IEC plug or a plug suitable for your local electrical supply If the plug has been removed or has to be removed the cable will contain...

Страница 5: ...symbol hot surfaces on the equipment Adopt suitable measures to prevent any danger arising from the formation of hot surfaces or electric sparks The pumps have no approval for operation in or for pum...

Страница 6: ...d substances might be released into the environment or into the pump housing To reduce the risk of leaks ask for a diaphragm pump with additional safety diaphragm Comply especially with notes on opera...

Страница 7: ...om mains and wait two minutes before starting mainte nance to allow the capacitors to discharge Before starting maintenance wait two minutes after isolating the equipment from mains to allow the capac...

Страница 8: ...o 5 8 o t 0 3 r e w o P W 0 8 1 0 0 2 z H 0 6 0 5 d e e p s d a o l o N n i m 1 0 0 8 1 0 0 5 1 t a t n e r r u c d e t a R V 0 0 1 z H 0 6 0 5 z H 0 6 V 0 2 1 z H 0 6 0 5 V 0 3 2 A A A 8 3 6 3 0 3 6...

Страница 9: ...a r e p o g n i r u d 5 8 o t 0 3 y l p p u s e g a t l o v f o e g n a r e l b i s s i m r e p m u m i x a M z H 0 6 0 5 0 1 V 0 3 2 o t 0 1 V 0 0 1 e s u F V 0 5 2 A 8 T m m 0 2 x 5 n o i t a r e p...

Страница 10: ...a l p r e v o c t o p h c t a C P P g n i t c e l l o c r e s n e d n o c r u o p a v e t s a w t s u a h x E k s a l f d e t a o c s s a l g e t a c i l i s o r o b s s a l g e t a c i l i s o r o B...

Страница 11: ...sk to collect condensate pump MZ 2C gas ballast valve PC 500 LAN controller VNC 1 vacuum connection VNC 1 connection pump mains connection VNC 1 on off switch diaphragm pump at the terminal box on off...

Страница 12: ...Attach the pipelines of the coolant circuit to the respective hose nozzles see im age at the waste vapour condenser Check hose connections prior to starting op eration of the cooling system Secure co...

Страница 13: ...ure after approx 15 minutes Prevent internal condensation transfer of liquids or dust The diaphragm and valves will be damaged if liquids are pumped in significant amounts Let the pump run with gas ba...

Страница 14: ...ent recovery rates at the exhaust waste vapour condenser Attention Notes concerning the operation of the exhaust waste vapour condenser Check hose connections prior to starting operation of the coolin...

Страница 15: ...ned catchpots Important Comply with regulations when disposing solvents condensates Reuse if possible purify if contaminated Shutdown Short term Has the pump been exposed to condensate Allow the pump...

Страница 16: ...number of version are displayed for 2s Display and keys key up select menu next menu item increase value key down select menu previous menu item decrease value key Enter start of control next step in...

Страница 17: ...is document with respect to his product Manual no 999084 05 10 2009 page 17 of 46 Connections serial interface RS 232 C vacuum connection connection mains cable 8 pin socket for connection of coolant...

Страница 18: ...connect the pressure trans ducer Condensate and deposits at the pressure transducer falsify the measurement result If residues occur or when working with aggressive or con densable substances install...

Страница 19: ...e Vacuum control pControl Controller controls pump and or isolation valve according to preset pressure in two point controller continuous pumping VARIO pump exactly according to preset pressure coolan...

Страница 20: ...eadjustment Adjusting the pressure transducer at atmospheric pressure and under vacuum see section Readjustment RS 232 parameters Setting of the interface parameters see section Interface parameters S...

Страница 21: ...tely only when the device menu is terminated Switch off the device to ignore a started new configuration of the controller 1013mbar Start Esc Esc VACUULAN V 2 00 Cal Sens Cal Atm see section Readjustm...

Страница 22: ...ase Upper pressure value p On factory set 200 mbar Condition for restart of the pump when the pressure is increasing If the actual pressure is higher than the upper set point the pump starts again Low...

Страница 23: ...local vacuum network VACUU LAN pSet 1 pSet 2 pSet 3 system pressure pressure in vacuum tubing network pumping unit Ports with non return valves Assembling the components of a VACUU LAN network A VACU...

Страница 24: ...plication Automatically controlled vacuum applications An increase of system pressure due to formation of gas e g pumping down and or programmed preset pressures can occur if several automatically con...

Страница 25: ...and laboratory Adjustment under vacuum Evacuate the controller via the vacuum connection to a pres sure 0 5 mbar e g by applying a good rotary vane pump Select program Cal Sens at controller The displ...

Страница 26: ...at any time Setting of the interface Setting of the interface parameters directly at the controller is described below Enter the device menu Select RS 232 The factory set values are underlined Baud 9...

Страница 27: ...ssure IN_PV_4 xxx min OnlyVACUU LAN Time of process IN_PV_5 XX XX h m OnlyVACUU LAN Preselected IN_CFG XXXXX at controller 0 Remote mode off 1 Remote mode on 0 not used 0 not used 0 No coolant valve 1...

Страница 28: ...mperature error 1 Error isolation valve 1 Error coolant valve 1 External errorr e g at VARIO Preset pressure IN_SP_1 XXXX mbar or Unit according to preselection XXXX Torr or XXXX hPa Status of IN_STAT...

Страница 29: ...controller 0001 to 1060 mbar hPa or 0001 to 0795Torr Pumping speed OUT_SP_2 XX X In Hz 01 0 to 60 0 99 9 corresponds HI Start up pressure OUT_SP_3 XXXX In unit according to selection at controller up...

Страница 30: ...nds Function Command Parameter Description The single abbreviations of a command are separated by underscores ASCII 5FH Command and parameter are separated by ignore characters ASCII 20H The string is...

Страница 31: ...nual flow control module VCL 01 677106 VACUU LAN shut off manual flow control moduleVCL 02 677107 VACUU LAN automatic control module VCL 10 677108 VACUU LAN manual flow control automatic control modul...

Страница 32: ...a r t e r u s s e r p e g a s s e m r o r r E s y S k c e h C s p e e b e v i f c i s a b n i s s e c o r p f o e m i T N A L U U C A V e d o m d e d e e c x e s k a e l r o f m e t s y s k c e h C e...

Страница 33: ...n e C e h t n i k a e l r o d e n o i t i s o p m u u c a v r o e n i l e p i p m e t s y s a h t i w p m u p k c e h C t a y l t c e r i d e g u a g m u u c a v k c e h c t r o p t e l n i p m u p e...

Страница 34: ...60 70 C Wait until temperature is reached Determine the vacuum level for a boiling point use published data of solvents Reduce pressure until a sufficient level of evaporation is attained If hysteresi...

Страница 35: ...e the pump and other compo nents from the vacuum system and the electrical supply Drain condensate if applica ble avoid the release of pollutants Separate pump from cooling water circuit Allow suffici...

Страница 36: ...the elbow fitting from the cover plate Through reassembly a leak may result Disconnect cables from the controller and unscrew controller from holding plate Unscrew socket head screw at rear of contro...

Страница 37: ...ump head and remove upper housing hous ing cover with housing cover insert head cover and valves Never remove parts by using a spiky or sharp edged tool e g screw driver we recommend to use a rubber m...

Страница 38: ...e pointer in the handle of theVACUUBRAND face wrench shows to the longer mark ing line Check diaphragm for damage and replace if necessary To do so use Phillips screw driver to remove four counter sun...

Страница 39: ...ney shaped opening beside valve By turning eccentric bushing bring connecting rod into up per turning point position Max stroke of the rod Pay attention that the diaphragm stays positioned centrally s...

Страница 40: ...ary recheck pump chamber Assembling the pump at the pump support and on the base plate Tighten union nuts first by hand and then tighten one full turn using open ended wrench w f 17 PC 600 LAN Assembl...

Страница 41: ...Cleaning and assembling components Cleaning the pressure transducer The controller itself is maintenance free If the vacuum system is contaminated oil particles etc contamination of the pressure tran...

Страница 42: ...the Federal Republic of Germany for the field of meteorology and certain sectors of safety engineering for the measurable variable pressure in the pressure range from 10 3 mbar to 1000 mbar in accorda...

Страница 43: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Страница 44: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil...

Страница 45: ...le dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne conception dessin et mod le vendu par nous m mes Cette d claration c...

Страница 46: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Отзывы: