Status of
IN_STAT_1 XXXXX
process control
Mode
: see IN_STAT
Status of operation : see IN_STAT
Isolation valve
0: closed
1: open
Cooling water
0: closed
valve
1: open
Relay
0: not operated
1: operated
Write commands
Function
Command
Parameter
Description
Preset pressure
OUT_SP_1
XXXX
In unit according to selection at controller
(0001 to 1060 mbar (hPa) or 0001 to 0795 Torr)
Pumping speed
OUT_SP_2
XX.X
In Hz (01.0 to 60.0, 99.9 corresponds HI)
Start-up pressure/ OUT_SP_3
XXXX
In unit according to selection at controller
upper pressure value
(see OUT_SP_1)
VACUU•LAN: start-up pressure ”
p On
”
Vacuum control: ”
p set”
+ ”
Δ
p
”
Switch-off time
OUT_SP_4
hh:mm
Hours:minutes
Lower pressure
OUT_SP_5
XXXX
In unit according to selection at controller
value
(see OUT_SP_1)
Vacuum control: upper pressure
Continuous pumping: lower pressure
Time delay
OUT_SP_6
hh:mm
Hours:minutes
(only VACUU•LAN;
otherwise process time)
Mode of operation OUT_MODE
X
0: VACUU•LAN
0S: VACUU•LAN blocked
1: Continuous pumping
2: Vacuum control
4: RC /RC 6 management
Read command
Function
Command Response
Description