Catchpot at inlet:
➨
Admit air or inert gas (via inlet of pumping unit) to atmospheric pressure. Remove
joint clip, remove catchpot and drain condensate.
➨
Reassemble drained catchpots.
Important:
Comply with regulations when disposing solvents/condensates. Reuse if
possible, purify if contaminated.
Shutdown:
Short-term:
Has the pump been exposed to condensate?
☞
Allow the pump to continue to run at atmospheric pressure for a few minutes.
Has the pump been exposed to media which may damage the pump materials or forms
deposits
?
☞
Check and clean pump heads if necessary.
Long-term:
➨
Take measures as described in section short-term shutdown.
➨
Separate pump from the apparatus.
➨
Drain catchpots.
➨
Close inlet and outlet port (e. g. with transport caps).
➨
Store the pump in dry conditions.