Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999084 / 05/10/2009
page 28 of 46
Read commands
Function
Command Response
Description
Error status
IN_ERR
XXXX
1: Last command to interface incorrect
1: Failure at pressure transducer
1: Overpressure
1: See ERR_1
Error status
IN_ERR_1
XXXX
1: Temperature error
1: Error isolation valve
1: Error coolant valve
1: External errorr (e.g. at VARIO)
Preset pressure
IN_SP_1
XXXX mbar or
Unit according to preselection
XXXX Torr or
XXXX hPa
Status of
IN_STAT
XXXX
process control
VACUU•LAN
0: Off
1: Process
2: Pressure<preset pressure
3: Pressure check
Continuous
0: Off
pumping
1: On
Vacuum control
0: Off
1: Pressure>upper pressure value
2: Preset pressure<pressure<
upper pressure value
3: Pressure<Preset pressure
4: End
RC 5
0: Off
1: Warmup
2: Pressrue>preset pressure
3: Pressure<preset pressure
Mode
0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5/RC 6 management
0: not used
Coolant valve
0: Closed
1: Open