1) Calibration must be made with new lamps only.
2) Lamps must be allowed to warm up for three minutes prior to any calibrations.
3) Open the enclosure door to expose the display plate. On the right side of the display plate, locate the
Run potentiometer, the Set potentiometer and the Cal switch.
4) First set the trip level. Push and hold the Cal switch and adjust the Set potentiometer until the display
reads 85%. When the UV intensity falls below 85% the system will go into the fault state. The UV unit is
designed to provide proper UV dosage at 85% of original UV intensity. With modern technology UV lamps,
this should occur at about 10,000 hours of continuous use (about one year).
5) Release the Cal switch and adjust the Run potentiometer until the display reads 100%.
6) When the board is in the normal state, test the calibration by adjusting the Run potentiometer until the
display reads below 85% to determine if the low UV lamp lights, indicating the board is functioning properly.
Readjust the Run potentiometer until the display reads 100%. (The three minute warm-up period will
restart). The UV Meter Board is now calibrated.
Содержание LEOPARD UV Series
Страница 2: ......
Страница 6: ......
Страница 35: ......
Страница 40: ...Replacement Parts List Electrical Enclosure Leopard Series Short Long Chambered Systems...
Страница 43: ...Model Number __________ Serial Number __________...
Страница 44: ...d Z s K D s s hs Y W K K Z K Z K t W W Z s s hs D Z s Z hs hs...
Страница 46: ...EKd...
Страница 47: ......