f) Sensor Probe Maintenance
1) Models with Standard UV Meter Board Monitor—
For models equipped with a standard UV Monitor, a sensor probe has been installed on the UV chamber.
The sensor probe contains an O-ring that will need to be replaced periodically. Annual routine maintenance
is recommended similar to lamp and quartz maintenance.
Unscrew the sensor probe from the sensor port. Inspect and clean as needed. A buildup can coat the
sensor probe and failure to clean may result in a false reading. It is recommended to replace the O-ring every
3-5 years as the O-ring can degrade over time.
Reattach sensor probe after routine maintenance has been performed.
2) Models with Standard UV Monitoring System or WatchDog Monitoring System—
a) Wetted Sensor Probe (SNSR0007 & SNSR0004)
For models equipped with either the Standard UV Monitoring System or WatchDog Monitoring System, a
sensor probe and sensor cable is used and attached to the UV chamber. The sensor probe consists of a an
O-ring that will need to be replaced periodically. Annual routine maintenance is recommended similar to
lamp and quartz maintenance.
You must shut down the system and depressurize the chamber before performing any maintenance.
Detach the sensor cable from sensor probe and unscrew the probe from the sensor port. Inspect
and clean as needed. A buildup can coat the sensor probe and failure to clean may result in a false
reading. It is recommended to replace the O-ring every 3-5 years as the O-ring can degrade over time.
Reattach sensor probe and sensor cable after routine maintenance has been performed.
b) Non-Wetted Sensor Probe with Viewing Window (SNSR0005)
For models equipped with the WatchDog monitoring system, a sensor probe has been installed on the UV
chamber. The sensor probe viewing window is made of a quartz disc that will need to be cleaned and
contains an O-ring that will need to be replaced periodically. Annual routine maintenance is recommended
similar to lamp and quartz maintenance. Before cleaning the viewing window, you must shut down the
system and depressurize chamber.
Unscrew the viewing window from the sensor port. Inspect and clean as needed. A buildup can coat the
quartz disc and failure to clean may result in a false reading. It is recommended to replace the O-ring every
3-5 years as the O-ring can degrade over time.
Reattach viewing window after maintenance has been performed.
sŝĞǁŝŶŐtŝŶĚŽǁ
WͬE^E^ZϬϬϬϱ
EŽŶ
-
tĞƩĞĚ^ĞŶƐŽƌWƌŽďĞ
^ƚĂŶĚĂƌĚhsDŽŶŝƚŽƌŝŶŐ^LJƐƚĞŵ
WͬE^E^ZϬϬϬϳ
tĞƩĞĚ^ĞŶƐŽƌWƌŽďĞ
tĂƚĐŚŽŐDŽŶŝƚŽƌŝŶŐ^LJƐƚĞŵ
WͬE^E^ZϬϬϬϰ
tĞƩĞĚ^ĞŶƐŽƌWƌŽďĞ
Содержание LEOPARD UV Series
Страница 2: ......
Страница 6: ......
Страница 35: ......
Страница 40: ...Replacement Parts List Electrical Enclosure Leopard Series Short Long Chambered Systems...
Страница 43: ...Model Number __________ Serial Number __________...
Страница 44: ...d Z s K D s s hs Y W K K Z K Z K t W W Z s s hs D Z s Z hs hs...
Страница 46: ...EKd...
Страница 47: ......